Temp. Deviation
The temperature difference between setting temperature and present
temperature is more than the range of the permission. The unit check it
once the present value reached to setting value. Once the gap is over the
permission range, the alarm is generated.
2
RPM Deviation
The rpm difference between setting rpm and present rpm is more than the
range of the permission. The unit check it once the present value reached
to setting value. Once the gap is over the permission range, the alarm is
generated.
3
Load Unbalance
Once the shaking system is vibrated by the present operating, the system
adjusts the shaking speed to the controllable stage. This is on the display.
4
Door Open
Once the door is opened, the temperature and rpm are not operated.
Once the door is opened, visual-audial alarm is generated. After 3 minutes
of door opening, the alarm is effective. Once the door is closed, alarm is
eliminated and the operating is back.
Alarm is not effective if Sound Off is set on the General Setting.
Deviation Alarm is not effective once the present value reaches to the setting value. (set time, in
permission range)
Deviation alarm is initialized under the below situation.
-
Set the setting value