18
Measuring method: Oscillometric
Operating method: Continuous operation
Pressure sensor: Resistive
Measuring range: DDIA: 15-160mmHg; SYS: 40-240mmHg
Pulse: 40 to 199 per minute
Cuff pressure display range: 0–295 mmHg
Memory: Automatically stores the last 120 measurements for 2 users (total 240)
Measuring resolution: 1 mmHg
Accuracy: Pressure within ± 3 mmHg / pulse ± 5 % of the reading
Power source: 2 AAA batteries, 1.5 V
Accessories: 13.5-19.5cm
Automatically power off : 60 seconds
Users: Adult
Software Version: V1.1.1
Expected service life of the device and accessories: 5 years
Technical alterations reserved!
10. EMC Declaration
1)
* This product needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided, and this unit can be affected by
portable and mobile RF communications equipment.
2)
* Do not use a mobile phone or other devices that emit electromagnetic fields, near the
unit. This may result in incorrect operation of the unit.
3)
* Caution: This unit has been thoroughly tested and inspected to assure proper
performance and operation!
4)
* Caution: this machine should not be used adjacent to or stacked with other equipment
and that if adjacent or stacked use is necessary, this machine should be observed to verify
normal operation in the configuration in which it will be used
Guidance and manufacture’s declaration – electromagnetic emission
The device is intended for use in the electromagnetic environment specified below. The customer of the user of the
device should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The
device
use RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.