V03
JIK-8 User Manual
47
【
ERROR MESSAGES
】
Error
Message
Reasons / Possible Caused
Solutions
E0
The CPU unable to read the
EEPROM
Contact the manufacturer or
nearest agent
E1
Unable to read the 3 points
calibration range
5HIHUWR³VHUYLFHPDQXDO´IRU
calibration procedures
E2
Zero Point is too high
(1) Make sure the pan is empty
when turn on the scale or
perform the 3 points
calibration.
(2) Check the connections of
wires
E3
Zero Point is too Low
(1) Make sure the pan is on the
scale or perform the 3
points calibration.
(2) Check the connections of
wires.
E4
Unstable zero point
(1) Make sure there is no winds
or vibration .
(2) Check the connections of
wires.
E5
(1)Load cell spec. not
compatible.
(2)calibrating weights mistake
(1) Replace with a compatible
load cell.
(2) Change with correct
calibrating weights.
E6
Load cell read out always the
same
(1)Check if load cell wire are
connected correctly.
E10
Optional RS-232(RTC)
batteries run out
Replace the batteries
E11
Unable to accumulate. Two
objects are with different units.
Press
twice to clear all
accumulation data or press
and return to normal weighing
mode.
Summary of Contents for JIK-8
Page 4: ......
Page 6: ......
Page 8: ......
Page 9: ......
Page 11: ......
Page 13: ......
Page 14: ......
Page 15: ......
Page 16: ......
Page 18: ......
Page 19: ......
Page 20: ......
Page 21: ......
Page 23: ......
Page 25: ......
Page 26: ......
Page 28: ......
Page 31: ......
Page 35: ......
Page 37: ......
Page 38: ......
Page 42: ......
Page 43: ......
Page 45: ......
Page 46: ......
Page 48: ......
Page 49: ......
Page 52: ......
Page 54: ......
Page 56: ......