JIK-8 User Manual
55
【
SINGLE POINT CALIBRATION FOR WEIGHT
】
Turn on the scale by holding down
key until CAP is shown on the lower left screen, i.e. the
scale has entered into the calibration mode.
If the calibration unit, capacity, and resolution have been set, you may skip Steps II~IV and
press
key to enter into Step V to perform zero point calibration.
You may select calibration unit (kg, g) by use of
key.
Press
or
key to shift the flickering digit left or right; press
,
key to set
any value between 1-9; after setting, press
key to enter into the next step.
Press
or
key to shift the flickering digit left or right; press
,
key to set
any value between 1, 2, 5; after setting, press
key to save and show the offset-value; press
WKHZHLJKLQJSDQJHQWO\LIWKHYDOXHFKDQJHVLW¶VQRUPDO
Press
key to perform zero point calibration; when CAL on the lower left stops flickering,
zero point calibration is completed with CAL **kg shown.
If the show value is very unstable, press
key to enter into stb adjustment function, use
key to extend the range of stb (it is recommended to adjust one segment each time),
after confirmation, press
key to save setting and the zero point calibration will be
performed automatically.
Or use
to skip the zero point calibration, and use last zero
point directly.
Summary of Contents for JIK-8
Page 4: ......
Page 6: ......
Page 8: ......
Page 9: ......
Page 11: ......
Page 13: ......
Page 14: ......
Page 15: ......
Page 16: ......
Page 18: ......
Page 19: ......
Page 20: ......
Page 21: ......
Page 23: ......
Page 25: ......
Page 26: ......
Page 28: ......
Page 31: ......
Page 35: ......
Page 37: ......
Page 38: ......
Page 42: ......
Page 43: ......
Page 45: ......
Page 46: ......
Page 48: ......
Page 49: ......
Page 52: ......
Page 54: ......
Page 56: ......