V03
JCAII
8
5. Press key and wait till
is displayed in the Total Count window.
6. Press key to save the settings and return to weighing mode
7-6 Set AD gain of remote platform
1. Under normal weighing mode, press for 3 seconds to enter the basic parameter setting.
2. Press
for 3 seconds to enter the advanced parameter setting.
3. Press key or to select parameter A-07
4. Press or to select the AD gain. (Options: 1/2/4/8/16/32/64)
5. Press to save and return to normal weighing mode.
7-7 Switch between Main scale and Remote platform
Press to switch between main scale mode and sub channel mode (remote platform)
Note
:
When accumulation is already in one channel, it can’t be accumulated in another channel.
When Hi/Lo checking is enable in one channel, Hi/Lo setting can’t be accessed and enable in
another channel.
8. Operation
8-1 Zero-point
Press if you do not see “
” but
symbol “
” appears
8-2 Tare & Preset Tare
8-2-1 Tare
1. Under the weighing mode, place a container weighing (eg. 0.100kg) on the pan.