PR-805/810/810
L
User Manual
102
Data Code
Description
117
Output Format
: <STAT>,<Aperture ID #>,<Aperture Name>,<Eff. BW> CRLF
Output Example
:
00000,0,2 deg,0.00
00000,1,1 deg,0.00
00000,2,0.2 deg,0.00
00000,3,0.1 deg,0.00
401
Output Format
: <STAT>,<# Stored Measurements in RAM> CRLF
Output Example
:
6
411
Output Format
: <STAT>
<Filename.mea>, <# Measurements> CRLF
Output Example
:
Display1.mea, 10
Display2.mea, 2
412
Output Format
: <STAT>
<Measurement ID #>, <Time Stamp> CRLF
Output Example
:
D412,TST.mea
00000
1,04-01-2008 19:14:04
2,04-01-2008 19:14:07
601
Output Format
: <STAT>,<PrimaryAcc>,<AddOn1>,<AddOn2>,<AddOn3>,
<Aperture ID #>, <Units>,<Gain>,<ND #>,
<Max Exp. Time>,<# of Cycles to Average>,<Freq. Response Filter>,
<MeasureShutter Status>,<AnalogOutput Status>
Output in numeric format/encoded.
Output Example
:
00000,0,-1,-1,-1,1,0,99,99,0,1,0,0,0
602
Output Format
: <STAT>,<PrimaryAcc>,<AddOn1>,<AddOn2>,<AddOn3>,
<Aperture ID #>, <Units>,<Gain>,<ND #>,
<Max Exp. Time>,<# of Cycles to Average>,<Freq. Response Filter>,
<MeasureShutter Status>,<AnalogOutput Status>
Output in numeric Verbose/Descriptive.
Output Example
:
00000,MS-75,None,None,None,1 deg,English,Auto Gain,Auto ND,0 msec,1 cycles,Open,
Closed after measurement,Analog Output Disabled