9. Product Specifications
9.1 Delivery form
– Programat S1
– Power cord
– Temperature Checking Set S1 Starter Kit
– Operating Instructions
– Sinter tray
– Sinter tray fork
– USB download cable
– Equipment Service Passport
– Various accessories
9.2 Technical data
Power supply
118–240 V / 50–60 Hz
Overvoltage category II
Contamination level 2
Acceptable voltage fluctuations
+/- 10%
Max. power consumption
16 A at 118 V
8 A at 240 V
Electrical fuses
250 V / T16 A high breaking (heating circuit)
Dimensions electrical fuses
Diameter 5 x 20 mm
Dimensions of the closed furnace
Depth: 430 mm / Width: 310 mm / 390 mm (with cooling tray)
Height: 570 mm
Usable size of the sintering chamber
Diameter 80 mm
Height 80 mm
Max. firing temperature
1530 °C/2786 °F
Weight
Furnace base: 10 kg
Furnace head: 17 kg
Safety information
The sinter furnace complies with the following guidelines:
– IEC 61010-1:2001
– EN 61010-1:2001
– UL 61010-1:2004
– CAN/CSA-C22.2 No. 61010-1:2004
– IEC 61010-2-010:2003
– EN 61010-2-010:2003
– CAN/CSA-C22.2 No. 61010-2-010:2004
Radio protections / electromagnetic compatibility
EMV tested
9.3 Acceptable operating conditions
Acceptable ambient temperature range:
+5°C to +40°C (+41°F to +104°F)
Acceptable humidity range:
80 % maximum relative humidity for temperatures up to 31°C
(87.8° F) gradually decreasing to 50 % relative humidity
at 40°C (104°F); condensation excluded.
Acceptable ambient pressure:
The furnace is tested for use at altitudes of
up to 2000 m above sea level.
9.4 Acceptable transportation and storage conditions
Acceptable temperature range
-20 to +65°C (-4°F to +149°F)
Acceptable humidity range
Max. 80% relative humidity
Acceptable ambient pressure
500 mbar to 1060 mbar
Use only original packaging of the Programat S1 together with the
respective foam material for shipping purposes.
30