Operating Instructions

Programat

®

P300

Summary of Contents for Programat P300

Page 1: ...Operating Instructions Programat P300...

Page 2: ...2...

Page 3: ...Explanation of the key functions 5 3 Program structure 5 4 Adjustable parameters and possible value ranges 5 5 Settings test programs and information 5 6 Explanation of the symbols on the display 5 7...

Page 4: ...hose connection 23 Rubber feet 24 Protective cover vacuum 25 Furnace head housing 26 Thermocouple plug 27 Plug fuse 28 Heater plug 29 Heater plug socket 30 Thermocouple plug socket 32 Leaf spring 33 A...

Page 5: ...5 40 25 34 6 10 35 23 47 33 1 7 5 2 3 4 8 9...

Page 6: ...6 37 39 36 11 38 25 19 13 12 49 26 30 29 50 51 27 28 11 32 43 12 13 15 19 46 21 17 18 16...

Page 7: ...y temperature 82 Closing time 83 Temperature increase 84 Holding temperature 85 Holding time 86 Vacuum on 87 Vacuum off 88 Long term cooling 90 Open furnace head 91 Close furnace head 92 Numeric keys...

Page 8: ...joy working with the P300 1 2 Introduction The signs and symbols in these Operating Instructions facilitate the finding of important points and have the following meanings Risks and dangers Important...

Page 9: ...must be observed The furnace must be operated under the indicated environmental and operating conditions Chapter 9 The Programat P300 must be properly maintained 2 1 1 Risks and dangers The furnace he...

Page 10: ...e skin grease contamination as the parts will be prematurely dama ged 2 1 12 Risk of crushing and burn hazard Never reach under the furnace head with the hand or other parts of the body during operati...

Page 11: ...nditions over an extended period of time Use only original spare parts The temperature range for faultless operation is 5 C to 40 C 41 F to 104 F If the furnace has been stored at very low temperature...

Page 12: ...uum may be created with a vacuum pump The firing process is controlled with the corresponding electronic con trols and a software Moreover the set and actual temperatures are continuously compared 3 4...

Page 13: ...ly use the two corresponding inserts Fold the side flaps 2 and combine the two packaging parts by means of the transportation flaps The packaging may be disposed with the regular household refuse 4 2...

Page 14: ...e 5 on the firing plate holder 48 Secure the cooling tray 34 with the two screws 35 including the silicone rest 47 Place the cooling tray 34 on the frame plate 7 Make sure that the cooling tray 34 is...

Page 15: ...ace head mounting 43 Ensure that the furnace head mounting mark 41 is aligned with the furnace base mounting mark 42 Make sure that the firing plate 5 is not damaged by mounting the furnace head Step...

Page 16: ...itional connections Power connection Please make sure that the voltage indicated on the rating plate complies with the local power supply Connect the power cord 16 with the power socket 17 of the furn...

Page 17: ...d 16 with the wall socket 2 Put the On Off switch 11 at the rear of the furnace on position I The furnace will now automatically conduct a self test The performance of all furnace components is automa...

Page 18: ...n menu The right key can be used to navigate between the settings and auxiliary programs Cursor keys 79 80 The current cursor position is indicated by means of a fix non blinking frame around the valu...

Page 19: ...uration and information 5 4 Adjustable parameters and possible value ranges Symbol Parameter Value range Value range P Program number P 1 100 B Stand by temperature 100 700 C 212 1292 F S Closing time...

Page 20: ...ntered by using the numeric keys The date can be selected by using the numeric keys Settings Write protection Ivoclar Vivadent optimized temperature control function Pre drying Vacuum test program Hea...

Page 21: ...reset to 0 Only after entry of the user code Settings Serial number Software version Furnace head firing hours Operating hours furnace Operating hours vacuum pump Latest start of calibration program...

Page 22: ...ox 10 seconds with the beeper signal set by the user If the STOP key is not pressed during this time the beeper sounds again after 5 minutes for a period of 5 minutes Subsequently there will be no fur...

Page 23: ...priate symbol first or second stage Two stage program stage 1 is shown stage 2 is shown If the cursor is positioned on the one stage symbol and the symbol is switched to the two stage symbol by pressi...

Page 24: ...ttom line until the next value is entered and successfully confirmed or the process is aborted with ESC The old valid value reappears Please refer to the parameter details for the value range 6 4 2 Pr...

Page 25: ...on the surface of the insulation are harmless and do not influence the function of the furnace in a negative fashion weekly Thermocouple 4 Check if the thermocouple is straight and in the right place...

Page 26: ...s the ENTER key The following is displayed c Insert silver wire into the Ivoclar Vivadent sample holder d Place the sample holder with the silver strip in the centre of the firing plate 6 e Press STAR...

Page 27: ...local Ivoclar Vivadent Service Center 13 System Heating muffle very old 23 The heating muffle is very old It is recommended to replace it After the error message has been acknowledged a firing program...

Page 28: ...ng program from the The program could not be continued since the USB memory stick is no longer present memory stick in progress memory stick no longer present 40 System Prolonged power failure 704 An...

Page 29: ...ure 69 System Reading calculating the 1204 no Error while measuring the furnace control temperature furnace control temperature 70 System Reading caluclating the 1205 Error while measuring the resista...

Page 30: ...and on again If it still does not work properly contact your local Ivoclar Vivadent Service Center Vacuum pump does not start working Is the vacuum pump fuse defective Check fuse and replace if necess...

Page 31: ...e Depth 368 mm width 303 mm 398 mm with Cooling tray Height 320 mm Usable size of the firing chamber Diameter 80 mm Height 48 mm Max firing temperature 1200 C 2192 F Weight Furnace base 12 3 kg Furnac...

Page 32: ...local Service Center Important information The current program table is also available at www ivoclarvivadent com The program tables can be downloaded from the Internet as PDF files Please make sure t...

Page 33: ...66 77 80 www ivoclarvivadent it Ivoclar Vivadent S A de C V Av Mazatl n No 61 Piso 2 Col Condesa 06170 M xico D F Mexico Tel 52 55 5062 1000 Fax 52 55 5062 1029 www ivoclarvivadent com mx Ivoclar Viv...