Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic environment - guidance
The “Bluephase N” uses RF energy only for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic equipment.
The “Bluephase N” is suitable for use in all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network that supplies buildings used for
domestic purposes.
P < 75W (pass without test)
P < 75W (pass without test)
Guidance and manufacturer‘s declaration - electromagnetic emission
The “Bluephase N” is intended for use in the electromagnetic environment specified below. The customer or the user of the “Bluephase N” should
assure that it is used in such an environment.
Guidance and manufacturer‘s declaration - electromagnetic immunity
The “Bluephase N” is intended for use in the electromagnetic environment specified below. The customer or the user of the “Bluephase N” should
assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line - line
± 2 kV line - earth
<5 % U
T
(>95 % dip in U
T
) for 0.5 cycle
40 % U
T
(60 % dip in U
T
) for 5 cycles
70 % U
T
(30 % dip in U
T
) for 25 cycles
<5 % U
T
(>95 % dip in U
T
) for 5 sec
3 A/m
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line - line
no prot. earth
<5 % U
T
(>95 % dip in U
T
) for 0.5 cycle
40 % U
T
(60 % dip in U
T
) for 5 cycles
70 % U
T
(30 % dip in U
T
) for 25 cycles
<5 % U
T
(>95 % dip in U
T
) for 5 sec
30 A/m
Electromagnetic environment - guidance
Floors should be concrete or ceramic tile. If floors are covered with synthetic material,
the relative humidity should be at least 30%.
Mains power quality should be that of typical commercial or dental environment.
Mains power quality should be that of typical commercial or dental environment.
Mains power quality should be that of typical commercial or dental environment. If the
user of the „Bluephase N“ requires continued operation during power mains interrup-
tions, it is recommended that the „Bluephase N“ be powered from an uninterruptible
power supply or battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in
a typical commercial or dental environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.