Warranty / Procedure
in Case of Repair
8. Product Specifications
The warranty period for Bluephase G4 is 3 years from
the date of purchase (battery: 1 year). Malfunctions
resulting from faulty material or manufacturing errors
are repaired free of charge during the warranty period.
The warranty however does not confer rights to
compensation for any material or non-material damage.
The apparatus must only be used for the intended
purposes. Any other uses are contraindicated. The
manufacturer does not accept any liability resulting
from misuse and warranty claims cannot be accepted in
such cases. This is particularly valid for:
• Damage resulting from improper handling, especially
incorrectly stored batteries (see Product Specifi cations:
Transportation and storage conditions).
• Damage to components resulting from wear under
standard operating conditions (e.g. battery).
• Damage resulting from external influences, e.g. blows,
drops to the floor.
• Damage resulting from incorrect set-up or installation.
• Damage resulting from connecting the unit to a
power supply, the voltage and frequency of which do
not comply with the ones stated on the rating plate.
• Damage resulting from improper repairs or
modifications that have not been carried out by
certified Service Centres.
In case of a claim under warranty, the complete
apparatus (handpiece, charging base, battery, power
cord and power pack) must be returned, carriage paid,
to the dealer or directly to Ivoclar Vivadent, together
with the purchase document. Use the original
packaging with the corresponding cardboard inserts for
transportation. Repair work may only be carried out by
a certified Ivoclar Vivadent Service Centre. In case of a
defect that cannot be rectified, please contact your
dealer or your local Service Centre (see addresses on the
reverse side). A clear description of the defect or the
conditions under which the defect occurred will
facilitate locating the problem. Please enclose this
description when returning the apparatus.
The product has been developed solely for use in dentistry. Processing should be carried out strictly
according to the Instructions for Use. Liability cannot be accepted for damages resulting from failure
to observe the Instructions or the stipulated area of application. The user is responsible for testing the
products for their suitability and use for any purpose not explicitly stated in the Instructions.
Light source
Ivoclar Vivadent Polywave
®
LED
Wavelength range
385–515 nm
Light intensity
High Power program: 1,200 mW/cm
2
± 10 %
PreCure program: 950 mW/cm
2
± 10 %
Operation
3 min on / 7 min off (intermittent)
Light guide
10 mm, autoclavable
Signal transmitter
Beeps every 10 seconds and every time the
Start/Stop button or the Time/Program
selection button is activated or the anti-glare
protection is enabled or the curing process is
aborted
Dimensions of the hand-
piece (without light
guide)
L = 170 mm, B = 30 mm, H = 30 mm
Weight of the handpiece
135 g (including battery and light guide)
Operating voltage of the
handpiece
3.7 VDC with battery
5 VDC with power pack
Operating voltage of the
charging base without
radiometer
100–240 VAC, 50–60 Hz, max 0.1 A
Operating voltage of the
charging base with
integrated radiometer
5 VDC
Power supply
Input: 100–240 VAC / 50 –60 Hz max 1 A
Output: 5 VDC / 3 A
Manufacturer: EDAC POWER ELEC.
Type: EM1024B2
Operating conditions
Tempe10 °C to +35 °C
Relative humidity 30 % to 75 %
Ambient pressure 700 hPa to 1060 hPa
Dimensions of the
charging base
D = 110 mm, H = 55 mm
Weight of the charging
base without radiometer
155 g
Weight of the charging
base with radiometer
145 g
Charging time
Approx. 2 hours (with the battery empty)
Power supply of the
handpiece
Li-Ionen battery
(approx. 20 min. with a new, fully charged
battery in the High Power program)
Transportation and
storage conditions
Temperature –20 °C to +60 °C
Relative humidity 10 % to 75 %
Ambient pressure 500 hPa to 1060 hPa
The curing light has to be stored in closed,
roofed rooms and must not be exposed to
severe jarring.
Li-ion battery:
– Do not store at temperatures above 40 °C /
104 °F (or 60 °C / 140 °F for a short period).
Recommended storage temperature
15–30 °C / 59–86 °F
– Keep the battery charged and store no
longer than 6 months.
Delivery form
1 Charging base with power cord and power
pack
1 Handpiece
1 Handpiece support
1 Light guide 10 mm
1 Anti-glare shield
3 Anti-glare cones
1 Pack of sleeves (1x 50 pcs)
1 Instructions for Use
Summary of Contents for Bluephase G4
Page 110: ...110 Bluephase G4 Bluephase G4 Ivoclar Vivadent...
Page 111: ...111 1 112 1 1 1 2 1 3 1 4 2 115 2 1 2 2 2 3 2 4 2 5 3 118 4 121 5 124 6 126 7 127 8 127...
Page 112: ...112 1 1 1 1 2 Click Cure 10 mm...
Page 113: ...113 1 3 High Power High Power 1 200 mW cm2 PreCure 2 950 mW cm2 20 15 10 Polyvision...
Page 114: ...114 1 4 2 4 20 High Power 3 High Power v x Click Cure polyvision...
Page 117: ...117 polyvision 4 Bluephase G4 515 nm Bluephase G4 Ivoclar Vivadent 40 C 60 C 6 10 20 20...
Page 118: ...118 Ivoclar Vivadent 3...
Page 119: ...119 Bluephase G4 20 2 2 5 114...
Page 120: ...120 Click Cure Bluephase G4 mW cm2 10 400 mW cm2 LOW...
Page 126: ...126 6 x x...
Page 146: ...146 Bluephase G4 Ivoclar Vivadent...
Page 147: ...147 1 148 1 1 1 2 1 3 1 4 2 151 2 1 2 2 2 3 2 4 2 5 3 154 4 157 5 160 6 162 7 163 8 163...
Page 148: ...148 1 1 1 1 2 Click Cure 10...
Page 149: ...149 1 3 High Power High Power 1200 2 PreCure 2 950 2 20 15 10...
Page 150: ...150 1 4 2 4 High Power 20 High Power 3 x Click Cure...
Page 153: ...153 4 Bluephase G4 515 Bluephase G4 Ivoclar Vivadent 40 C 60 C 6 10 20 20...
Page 154: ...154 Ivoclar Vivadent 3...
Page 155: ...155 Bluephase G4 20 2 2 5 150...
Page 156: ...156 Click Cure Bluephase G4 2 10 400 2 LOW...
Page 162: ...162 6 x x...