14
FEATURES AND OPTIONS
INFORMA
TION DISPLA
Y
1
FIX ME
2
FIX ME
3
FIX ME
4
FIX ME
5
FIX ME
FIX ME CAL
ORAL
FIX ME
BROKEN PROBE
RECTAL
FIX ME
BROKEN PROBE
• Calculated ROM CRC does not match copy burned into ROM.
Requires Service.
• RAM test failed. Requires Service.
• Analog fault detected. A/D reference voltages do not match
expected /- tolerances. Requires Service.
• Heater Interlock fault detected. Heater interlock has failed to
disable the probe heater circuitry. Requires Service.
• The instrument has detected an internal malfunction.
Requires service.
• A/D calibration phase timed out. “
CAL
” is displayed in LCD
panel. Requires Service.
• The instrument has detected a broken oral (blue) probe.
Requires Service.
• The instrument has detected a broken rectal (red) probe.
Requires Service.
SOLUTION
PROBLEMS/SOLUTIONS
DISPLAY MESSAGE