7
Configuration
ENT
ENT
pH
CAL
CONTROL
ALARM
REGISTER
SP
07.00
4.2 Setpoint configuration
Setting the desired values for pH for the two control channels by reference values or setpoints for alarm.
To enter the setpoint:
change the value with the up/down
arrows and confirm with ENT
4.3 Calibration
Calibration of the sensors used for pH and flow rate. The calibration menu is in each parameter menu.
4.3.1 pH calibration
ENT
ENT
pH
CONTROL
ALARM
REGISTER
SP
07.00
PH CAL.
Point 1
07.00
Point 2 04.00
Reading 07.23
PH CAL.
Point 1
07.00
Point 2 04.00
Reading 07.23
If necessary for pH calibration point 1, modify the buffer value using the up/down arrows. Insert the sensor
into the point 1 buffer, wait until the sensor reading is stable and confirm with ENTER.
Repeat the operation with the point 2 buffer. Before putting the sensor in the point 2 buffer, clean the
sensor with distilled water and dry it well (with tissue paper) to avoid contamination of the buffer. If at any
time during the calibration process you are not sure of the calibration, you can exit without confirming by
pressing ESCAPE.
CAL
4.3.2 Flow rate calibration
ENT
m3h
0.20 %
SP
CAL
CONTROL
ALARM
ALARM DETECTOR
REGISTER
Q CAL
.
T
ype:
Low Freq
K: 5.00
l/p
Ti
me Q=0: 015
s
Select the type of flow meter:
- Low Freq:
Low frequency flow meter. Flow meters with pulse frequencies of between a minimum of 1 pulse every
200 seconds and a maximum of 30 pulses per second (30 Hz).