22
Specifying Continuous POCT-1A Communication
Note
: Communication is initiated from the instrument only when the CLSI POCT-1A compliant
communication protocol is used. If the ITC proprietary communication protocol is used,
communication is initiated from a personal computer or network.
1.
From the Supervisor menu, press
7
to display the Communications menu:
2.
Press
3
to select the desired communication option when the CLSI POCT-1A compliant
communication protocol is used. (Display
OFF
to initiate communication only if requested, or
display
ON
to initiate continuous communication whenever the instrument is started.)
Note
: Specifying the communication option acts as a toggle. If initiation of communication upon
startup is already specified (
Push POCT
ON
is displayed), this can be cancelled by pressing
3
again
to display
Push POCT OFF
.
Entering Network Communication Parameters
Some or all of the following parameters may need to be entered in order for the instrument to
communicate over a network:
•
IP Address
•
Gateway
•
NetMask
•
Remote IP
•
Remote Port
Note
: Consult with the network administrator for your facility to determine which parameters
must be entered and what those settings should be.
1.
From the Supervisor menu, press
7
to display the Communications menu:
2.
(If the CLSI POCT-1A compliant communication protocol is currently selected): Press
1
to toggle the
communication protocol to the ITC proprietary communication protocol:
Note
: The CLSI POCT-1A compliant communication protocol must be
OFF
when running XPORT
Setup.
3.
Press
4
to start the XPORT Setup module. The message
XPORT Setup Please Wait
is displayed,
followed by the first page of the XPORT Setup module:
Note
: If the message
XPORT Setup NOT RESPONDING!
is displayed, contact ITC Technical Support
at (800) 631-5945 or (732) 548-5700, or by e-mail at [email protected].