16
4.
Enter the new time using the numeric keys.
Note
: Use the 24-hour format (i.e., enter 15:30 for 3:30 pm). The leading zero is required for
single digit times (i.e., enter 06:30 for 6:30 am).
5.
Press and hold down
ENTER
.
Stored
is displayed and the entry is saved.
6.
Press
CANCEL
to return to normal operation.
To Set the System Date:
1.
Press and hold
START
to turn on the instrument.
2.
Press
0
(zero) to display the main menu.
3.
Press
3
to display the date:
4.
Type the new date using the numeric keys.
Note
: Use the US (i.e.,
MM/DD/YY
) or the IVDD (i.e.,
YY-MM-DD
) date format. The leading zero
is required for single digit dates (i.e., enter 06/01/06 for June 1, 2006).
5.
Press and hold down
ENTER
.
Stored
is displayed and the entry is saved.
6.
Press
CANCEL
to return to normal operation.
Instrument Shutdown
1.
Press
0
(zero) to display the main menu:
2.
Press
4
. The instrument is shut down.
Note
: The instrument can also be shut down by pressing and holding down the
START
key for
four seconds. The instrument will automatically shut down after being idle for five minutes if
there is no cuvette inserted in the test chamber and the instrument is running on the battery.
Displaying System Information
The instrument software version, instrument serial number, and other system software information for the
HEMOCHRON
Signature Elite
instrument can be displayed.
1.
Press and hold
START
to turn on the instrument.
2.
Press
0
(zero) to display the main menu:
3.
Press
5
to display the system information:
(example only)
Note
: The Main Menu will again be displayed after several seconds, or press
CANCEL
to return
to normal operation.
Summary of Contents for HEMOCHRON Signature Elite
Page 81: ...81 This page intentionally left blank...
Page 84: ...84 HX1101 0310...