The functions on"7.1" page 35 and POS. "7.2" page 35, if enabled and set on pulse value, allow
to set a defined amount of volume to get one pulse.
To set pulse value set the 2 fields, from left to right: 1)unit of measure, 2) numeric value.
The selection is performed by positioning the cursor in the field that has to be modified.
To change the unit type (metric, imperial) see menu 2 units.
The value of Pls1-2 depends on nominal diameter (max. flow rate) units.
The available units are described in the function "1.4" page 33.
When the pulse value is defined, also the duration of the pulse shall be set. It’s expressed in milliseconds and the
range is from 0.4 and 1250 ms. Be carefull to set a right value of pulse duration, according to the receiver. It shall
be compatible with the external device which receives the pulses.
ATTENTION:
If the pulse is too long receiver’s coils may burn out or if the pulse is too short, the counter may not
be able to recognize the pulses.
(POS. 3.3-3.5) OutPut 1-2 Pulse Value
[Pls1-2= dm³ x.xxxxx]
AL2
[OP1PV-OP2PV]
(POS. 3.4-3.6) OutPut1-2 Pulse Time
[Tpls1-2= ms x.xxxxx]
AL2
[OP1PT OP2PT]
(POS. 3.7-3.8) OUtput 1-2 Full scale Frequency
[Frq1-2= x.xxxxx]
AL2
[OU1FF-OU2FF]
Frq1 and Frq2 are activated with POS. 7.1 see page 35
POS. 7.2 see page 35 enabled and set to the value
freq+/-/+-. Setting duration of the pulse generated on channel 1 and 2.