MENU 11 - FUNCTION
The following functions are activated by first pressing the “ENTER” and then the “ESC” when the screen appears
“confirm” to start the function.
Reset total direct totalizer for direct flow rate (+)
R
eset total partial totalizer for direct flow rate (+)
Reset total reverse totalizer for direct flow rate (-)
Reset partial reverse totalizer for direct flow rate (-)
This function resets the parameters of the sensor factory default. To Load the saved files see function (11.7).
This function resets the parameters of the converter factory default. o Load the saved files see function (11.8).
This function save the parameters of the sensor factory default.
This function loads the data from a converter to another.
Perform manually a board’s calibration. Press Enter and the message “ EXECUTE?” will be visualized on the display
then press long the key Enter to proceed. Press any other key to delete the operation.
If the sensor table is valid, the calibration is performed also when one of the following paramenter has been change:
1.
SENSOR DIAMETER -> Menu Sensor1
2.
SENSOR MODEL -> Menu Sensor1
3.
Exc. CURRENT -> Menu Sensor1
4.
S. Freq. -> Menu Sensor1
To check the calibration status, active or inactive, type the command MCP Calic? and check as follows:
CALIC = 1 calibration in progress
CALIC = 0 calibration terminated
(POS. 11.1) Volume Totalizer Total Positive Reset
[T+ RESET= ON]
AL3
[VTTPR]
(POS. 11.2) Volume Totalizer Partial Positive Reset
[P+ RESET= ON]
AL3
[VTPPR]
(POS. 11.3) Volume Totalizer Total Negative Reset
[T- RESET= ON]
AL3
[VTTNR]
(POS. 11.4) Volume Totalizer Partial Negative Reset
[P- RESET= ON]
AL3
[VTPNR]
(POS. 11.5) Load Factory Default Sensor Data
[Load sens.f.def= ON]
AL3
[LFDSD]
(POS. 11.6) Load Factory Default Converter Data
[Load conv.f.def= ON]
AL3
[LFDCD]
(POS. 11.7) Save Factory Default Sensor Data
[Save sens.f.def= ON]
AL6
[SFDSD]
(POS. 11.8) Save Factory Default Converter Data
[Save conv.f.def= ON]
AL6
[SFDCD]
(POS. 11.9) CALibration Immediate Command
[Calibration]
AL5
[CALIC]
Save conv.f.def= ON. "meter data" page 76
Sensor ReFerence Data Save
[MCP ONLY]
AL4
[SRFDS]
MENU 11 - Function: ONLY MCP FUNCTIONS