22
1
2
180°
Replacing lamps
The replacement of the lamps must be carried out according to the working hours due to the decrease in the efficiency of the
same. The change is required every 9000 working hours which correspond to approximately one year in continuous service.
Before starting this type of maintenance, make sure that the equipment is disconnected from the power supply.
To access the lamps, simply remove the protective gray fixed by means of 4 blind nuts.
Once the lamp is accessed it can be easily replaced by rotating it 180 ° before removing it from the support feet, as in the
case of a common neon tube.
Subsequently, the new lamps are inserted by doing the opposite.
In order to maintain the sanitizing and germicidal efficiency of the Stril Room components, it is essential to scrupulously
comply with the maintenance program indicated below.
CAUTION!! Before any type of maintenance, especially internal ones, it is mandatory to disconnect the
components from the power supply.
For some of the maintenance operations, it is necessary to refer to the working hours of the lamps which
are stored by means of the hour meter located inside the control console.
Recommended maintenance schedule
Operation description
working time / hours
Cleaning of the reflecting parabolas
At least once a month in order to keep the reflective walls
clean for maximum efficiency
Change of lamps + lamp holders
every 9000 hours (about one year in continuous service)
Replacement of the IR sensor
every 9000 hours and in any case depending on the state of
wear of the external lens which could compromise correct
presence detection
MAINTENANCE