46
[email protected] | +49 921 150845-0
maximum height of 12 cm, a minimum width of 88 cm and a minimum length of 197 cm (mattress and
foot block together).
For safety reasons, a distance of 22 cm must be maintained between the upper edge of the
mattress (unloaded) and the upper edge of the uppermost wooden side rails (side rails in
upper position).
Mattresses with high volume weights are only permitted if the weight of the mattress and
the patient combined does not exceed the safe working load of the bed.
7
Ambient conditions
According to DIN EN 60601-2-52, the medical device can be used in the following application
environment:
Application environment 3:
Long-term care in a medical setting where medical supervision is required and monitoring is provided
as necessary; an ME device may be provided for a medical procedure to maintain, improve, or support
the patient's condition.
Application environment 4:
ME device to alleviate or compensate for an injury, disability or illness in home care.
A maximum noise level of 49 dB (A) occurs during adjustment of the electric drives.
7.1
Storage conditions
Storage conditions
min. -10 °C max. +50 °C
Relative humidity
min. 20 % max. 80 %
Air pressure (at altitude
≤
3000
m)
min. 700 hPa max. 1060 hPa
Table 3: Storage conditions
!
!