ITA Series – Table Top Precision Weighing Balance

ITA Series – Table Top Precision Weighing Balance

5.5

AUTO ZERO SETTING: Press the 
MODE  button  and  then  switch 
on  the  scale.  The  display  will 
show A0 – X (0, 1, 2, 3).This is the 
auto zero setting. You can select 
the value using the UP button. 3 means 3d in d mode and 3e in e 
mode (remember d =e/10 and e is the scale interval. That is e = 
5gms and d = 0.5gms) similarly for 3, in d=e/10 mode it will auto 
zero upto 1.5gms, and in e mode it will auto zero upto 15 gms. 
Default setting of Auto-Zero mode is 3d. Password to be entered.

5.6

MODE  :  This  setting  sets  the  valid  con  version  units  /  modes 
available  in the  system. The  available  units are  carat,  tola, piece 
counting mode. Selecting all the modes to YES will allow you to see 
the  weight  kept  on  the  pan  in  different  modes  by  pressing  the 
MODE button. Remember setting any mode to NO will disable the 
scale to weigh in that particular mode.

ALL THE ABOVE SETTINGS ARE AVAILABLE WHEN DOING THE 
INTERNAL  CALIBRATION  OF  THE  SCALE.  THE  METHOD  FOR 
INTERNAL CALIBRATION IS DESCRIBED BELOW. THIS PROCESS 
MUST BE DONE BY AUTHORISED SERVICE ENGINEERS ONLY.

6. CALIBRATION

 

All  ISHTAA  weighing  scales  come  pre-calibrated  from  the  factory. 
However  the  scale  may  require  re-calibration.  Temperature  changes, 
geographic  gravity  variations,  altitude  changes  and  abuse  are  few 
reasons why a scale may need recalibration. The scales can be calibrated 
using a known external weight.

Switch  off  the  scale.  Press  the  MODE 
button and switch on the scale. You will 
see  PASS.  Press  MODE  again  and  you 
will  see  “0000”  on  the  display.  The 
password is 1072 and is to be entered 
as follows 
a. Press the UP key once. 1 time only - 
The display “0000” will blink but remain 
as it is.
b. Press the SHIFT key 2 times. Display 
will be same that is “0000”.
c. Press the UP key 7 times. The display 
“0000” will blink but remain as it is.
d.  Press  the  SHIFT  key  1  time.  The 
display will be same that is “0000”.
e. Press the UP key 2 times. The display 
“0000” will blink but remain as it is.
f. Press the MODE key and you will see 
CAP on the display.

You  will  now  see  “CAP”  on  the  scale. 
Press Mode button and you can select 
the CAP option described in section 5.1 
as “YES/NO”. Default option is Yes. You 
can toggle the setting by pressing the 
UP  button.  Press  the  Mode  button  to 
select the desired setting.

You  will  now  see  “buz”  on  the  scale. 
Press Mode button and you can select 
the BUZ option described in section 5.2 
as “YES/NO”. Default option is Yes. You 
can toggle the setting by pressing the 
UP  button.  Press  the  Mode  button  to 
select the desired setting.

STEP

DISPLAY

PROCEDURE

6.1 INTERNAL SERVICE CALIBRATION 

 

 

 

1.

2.

3.

6

7

Summary of Contents for ITA-310-G

Page 1: ...Scale Users Guide ISHTAA SCALES INC INDIA www ishtaascales com ITA Series Table Top Precision Weighing Balance ITA Series G Precision Table Top Weighing Balance TM ...

Page 2: ...nce ITA Series Table Top Precision Weighing Balance Contents 1 Technical Data 1 2 Installation 2 3 Controls Functions 3 4 User Operation 4 5 Scale Settings 5 6 Calibration 7 7 Error Codes 11 8 Trouble Shooting 11 9 Warranty 12 GS ISI 13 04 09 1T ...

Page 3: ...CATE Customer Name ______________________________________________ _____________________________________________________________ _____________________________________________________________ Phone _______________________ Mobile _________________________ Date of Purchase ______________ Invoice DC No __________________ Model ______________________ Sl No __________________________ Dealers Agent Name _...

Page 4: ... Unpack and verify that the following components have been included a ITA Series G Precision Balance b Instruction Manual c AC Adapter Before you install the scale identify the best location for the equipment The proper environment enhances its operation and longevity Keep in mind the following factors which might have a negative influence on the scale s operation a Vibration Vibration diminishes ...

Page 5: ... selected setting and exit from setting mode External Calibration of Scale STEP DISPLAY PROCEDURE 1 2 3 4 5 6 Press and hold the tare button while switching on the scale You will see CAL 0 on the display for a second and then some counts will be displayed Press the mode button once you will see Load on the display Keep a known weight on the pan for eg 200 gm and press mode button The display will ...

Page 6: ...SCALE ON OFF The scale is switched ON OFF by using the Power switch located on the right side of the scale Note Make sure the weighing pan is empty before turning on the scale 4 2 TARE ZERO The machine has a common Tare Zero button TARE ZERO BUTTON This button is used for taring a weight to display 0 000 Place a container on the weighing pan say 200g then press the TARE button The weight is zeroed...

Page 7: ...ight most digit of the display and if the DSP 0 setting is set to NO the display will show all the 4 zeros on the display It is important to know that DSP 0 mode if set to YES increases the battery back up time STEP DISPLAY PROCEDURE 4 5 6 7 8 9 You will now see P off on the scale Press Mode button and you can select the P OFF option described in section 5 3 as YES NO Default option is Yes You can...

Page 8: ...r the scale may require re calibration Temperature changes geographic gravity variations altitude changes and abuse are few reasons why a scale may need recalibration The scales can be calibrated using a known external weight Switch off the scale Press the MODE button and switch on the scale You will see PASS Press MODE again and you will see 0000 on the display The password is 1072 and is to be e...