background image

STEP

DISPLAY

PROCEDURE

10.

11.

12.

If  you  put  200gm  weight,  Enter 
“200.000”. Change the rightmost digit 
by  pressing  the  UP  button.  Shift  the 
changed  digit  to  left  by  pressing  the 
“shift” button once. Enter the value as 
shown using the key combination.

Press the mode button once again now 
you will see “CAL- F” for one second and 
you will see the full load counts. 

Press the mode button and you should 
see  “UNLOAD”  on  the  display.  Lift  the 
weight from the pan and press the mode 
button  to  finish  the  calibration.  The 
m a c h i n e   s h o u l d   r e s t a r t   n o w  
automatically.

 

 

 

 

ITA Series – Table Top Precision Weighing Balance

ITA Series – Table Top Precision Weighing Balance

4 USER OPERATION

4.1

SWITCHING SCALE - ON / OFF : 

The scale is switched ON / OFF by 

using  the  Power  switch  located  on  the  right  side  of  the  scale.   

Note :

  Make sure the weighing pan is empty before turning on the 

scale.

4.2

TARE / ZERO : 

The machine has a common Tare/Zero button. TARE 

/ ZERO BUTTON : This button is used for taring a weight to display 
0.000.  Place  a  container  on  the  weighing  pan  (say  200g),  then 
press the TARE button. The weight is zeroed and the machine will 
display zero.

4.3

ACCUMULATION : 

Keep a weight on the weighing pan(ex 200g). 

The display will show the weight. Now press UP. The display will 
show M PLUS. Now keep another weight (ex 300g), the display will 
show its weight. Now press UP button. To recall the weight press 
the shift button.

4.4

MODE :

 This button is used to select the different units of weight 

available in the scale, that is to choose between the units of carat, 
tola, piece counting and gram.

 

4

9

Summary of Contents for ITA-310-G

Page 1: ...Scale Users Guide ISHTAA SCALES INC INDIA www ishtaascales com ITA Series Table Top Precision Weighing Balance ITA Series G Precision Table Top Weighing Balance TM ...

Page 2: ...nce ITA Series Table Top Precision Weighing Balance Contents 1 Technical Data 1 2 Installation 2 3 Controls Functions 3 4 User Operation 4 5 Scale Settings 5 6 Calibration 7 7 Error Codes 11 8 Trouble Shooting 11 9 Warranty 12 GS ISI 13 04 09 1T ...

Page 3: ...CATE Customer Name ______________________________________________ _____________________________________________________________ _____________________________________________________________ Phone _______________________ Mobile _________________________ Date of Purchase ______________ Invoice DC No __________________ Model ______________________ Sl No __________________________ Dealers Agent Name _...

Page 4: ... Unpack and verify that the following components have been included a ITA Series G Precision Balance b Instruction Manual c AC Adapter Before you install the scale identify the best location for the equipment The proper environment enhances its operation and longevity Keep in mind the following factors which might have a negative influence on the scale s operation a Vibration Vibration diminishes ...

Page 5: ... selected setting and exit from setting mode External Calibration of Scale STEP DISPLAY PROCEDURE 1 2 3 4 5 6 Press and hold the tare button while switching on the scale You will see CAL 0 on the display for a second and then some counts will be displayed Press the mode button once you will see Load on the display Keep a known weight on the pan for eg 200 gm and press mode button The display will ...

Page 6: ...SCALE ON OFF The scale is switched ON OFF by using the Power switch located on the right side of the scale Note Make sure the weighing pan is empty before turning on the scale 4 2 TARE ZERO The machine has a common Tare Zero button TARE ZERO BUTTON This button is used for taring a weight to display 0 000 Place a container on the weighing pan say 200g then press the TARE button The weight is zeroed...

Page 7: ...ight most digit of the display and if the DSP 0 setting is set to NO the display will show all the 4 zeros on the display It is important to know that DSP 0 mode if set to YES increases the battery back up time STEP DISPLAY PROCEDURE 4 5 6 7 8 9 You will now see P off on the scale Press Mode button and you can select the P OFF option described in section 5 3 as YES NO Default option is Yes You can...

Page 8: ...r the scale may require re calibration Temperature changes geographic gravity variations altitude changes and abuse are few reasons why a scale may need recalibration The scales can be calibrated using a known external weight Switch off the scale Press the MODE button and switch on the scale You will see PASS Press MODE again and you will see 0000 on the display The password is 1072 and is to be e...