ITA Series – Table Top Precision Weighing Balance
ITA Series – Table Top Precision Weighing Balance
5
SCALE SETTINGS
5.1
CAP – This setting enables or
disables the capacity display during
startup of the scale. The valid values
are YES / NO.
5.2
BUZ: This setting enables or disables
the buzzer sound on stability of
weight. When you keep or remove
weight on the machine, once the
weight on the machine stabilizes,
the beep sound is generated. Valid settings are YES / NO.
5.3
P-OFF: This setting can enable or
disable the sleep mode. When the P-
OFF setting is set to YES, if the scale
is idle for two minutes, that is there
is no weighing being done,
the display will simply show the
decimal point of the right most digit on the display. If machine is in idle
condition, the scale will automatically go to SLEEP mode to increase
the battery backup time. If you keep the weight, the scale will
automatically come out of the SLEEP mode and shows the weight. The
Normal backup time when the sleep mode is switched off is 30 Hours,
and with the SLEEP mode switched on, the battery back-up time
increases to more than 60 Hrs.
If the USER wants to switch off the SLEEP MODE simply set the P-OFF
setting to NO.
5.4
DSP-0 : This setting sets the no of
zeros to be shown on the display.
When the DSP-0 setting is set to
YES, if there is no weight on the pan, that is there is no weighing been
done, the display will simply show a single zero of the right most digit
of the display and if the DSP-0 setting is set to NO, the display will
show all the 4 zeros on the display. It is important to know that DSP-0
mode if set to YES increases the battery back up time.
STEP
DISPLAY
PROCEDURE
4.
5.
6.
7.
8.
9.
You will now see “P-off” on the scale.
Press Mode button and you can select the
P-OFF option described in section 5.3 as
“YES/NO”. Default option is Yes. You can
toggle the setting by pressing the UP
button. Press the Mode button to select
the desired setting.
You will now see “DSP 0” on the scale.
Press Mode button and you can select the
DSP0 option described in section 5.4 as
“YES/NO”. Default option is Yes. You can
toggle the setting by pressing the UP
button. Press the Mode button to select
the desired setting.
You will now see “AUTO-0” on the scale.
Press Mode button and you can select the
AUTO-0 option described in section 5.5.
Default option is 3.0d. You can toggle the
setting by pressing the UP button. Press
the Mode button to select the desired
setting.
You will now see “MODE” on the scale.
Press mode button and you can select the
MODE option described in section 5.6.
Default option for all the modes is YES .
You can toggle the setting by pressing the
UP button. Press the MODE button to
select the desired setting.
You will now see “CAL 1” on the display.
press mode button now you will see “CAL
0” on the display for a second. and the
display will show some internal counts.
Press mode button and you will see
“LOAD” on the display. Place a known
weight on the pan eg. 200g. Press the
mode button and you will see “0000” on
the display or the previous calibration
weight value.
8
5