8
Status register 2
(BA + 0x30)
Bit
Name
Description
Remark
D15
REG2ER_2
Quality of output voltage not given at present
D14
REG1ER_2
V
max
or I
max
is / was exceeded
Channel
D13
EXTINH_2
External inhibit was / is active
D12
RANGE_2
V
set
to V
max
ratio > 1
D(BA+0x08) > V
max
B
D11
KEY_CHANGED
A frontpanel switch position was changed
ON_OFF_2,
IN_EXT_2, KILL_2
D10
EOP_2 V
out
has reached set value
End of process_2
D9
ILIM_2 I
out
was > I
max
programmable
Current trip
D8
D7
REG2ER_1
Quality of output voltage not given at present
D6
REG1ER_1
V
max
or I
max
is / was exceeded
Channel
D5
EXTINH_1
External inhibit was / is active
D4
RANGE_1
V
set
to V
max
ratio > 1
A(BA+0x04) > V
max
A
D3
KEY_CHANGED
A frontpanel switch position was changed
ON_OFF_1,
IN_EXT_1, KILL_1
D2
EOP_1 V
out
has reached set value
End of process_1
D1
ILIM_1 I
out
was > I
max
programmable
Current trip
D0
TOT
Timeout error
New initialisation
The individual bits are set on the occurrence of the event. A general clear is performed after readout.
If the Output voltage was permanently switched off by exceeding V
max
or I
max
(ENABLE KILL resp. Current trip),
or INHIBIT respectively, the error bits (REG1ER_, EXTINH_, ILIM_) have to be reset by reading
”
Status
register 2
”
before an output voltage can be set again.
Start voltage change
(BA + 0x34 / Ba + 0x38)
A change in the output voltage of set voltage (BA + 0x04 / BA + 0x08) is performed by reading these registers.
Writing to the registers stores the data as new set voltage (V
set
≤
V
max
) and starts the voltage change.
The change of output voltage is blocked, if the conditions are unavailable to start voltage change corresponding
these description.
Command execution can be checked by reading status register 1 (BA + 0x00). The bits D14 (channel B), D6
(channel A) respectively are set on start of voltage change. Actual voltage reaching the set voltage is flagged by
the bits D10, D2 of status register 2 (BA + 0x30) respectively. An interruption of the voltage change (e.g.
external INHIBIT is active) is also ascertainable.
Module identifier
(BA + 0x3C)
D15 .. D0
4 digit serial number, BCD coded
Current software limit
(BA + 0x44 / BA + 0x48)
The maximal output current per channel corresponding current resolution. If the output current exceeds the
programmable limit, the output voltage will be shut off permanently by the software (Current trip). The Current
trip reaction time is lessen to 60 ms.
Writing to the registers with maximal output current = 0, not a current trip will be programmed.
Hardware maximum output current limit (I
max
) worked independently of programmable current software limit.