Status register 1
(BA + 0x00)
Channel Bit Name
Description
0
1
D15
ERROR_2
Error on Channel B
Channel ok
Error
D14
STATV_2
Status
V
Out
V
out
stable
V
out
in change
D13
TRENDV_2
Ramp up / down
V
out
falling
V
out
rising
B
D12
KILL_2
KILL switch setting
Disabled
Enabled
D11
ON_OFF_2
HV-ON/OFF switch setting
On
Off
D10
POL_2
Polarity
V
Out
Negative Positive
D9
IN_EX_2
CONTROL switch setting
DAC
Manual
D8
VZ_2
V
Out
= 0
V
out
<>0 V
out
=0
D7
ERROR_1
Error on Channel A
Channel ok
Error
D6
STATV_1
Status
V
Out
V
out
stable
V
out
in change
D5
TRENDV_1
Ramp up / down
V
out
falling
V
out
rising
A
D4
KILL_1
KILL switch setting
Disabled
Enabled
D3
ON_OFF_1
HV-ON/OFF switch setting
On
Off
D2
POL_1
Polarity
V
Out
Negative Positive
D1
IN_EX_1
CONTROL switch setting
DAC
Manual
D0
VZ_1
V
Out
= 0
V
out
<>0 V
out
=0
This register is representing the general status of the VHQ.
”
Error“ is formed by the logic or of REG2ER_, REG1ER_, EXTINH_, RANGE_ and ILIM_
from
”
Status register 2
”
.
”
V
out
=0
”
is formed by DAC output = 0 and actual voltage < 5 V.
Set voltage Channel A/B
(BA + 0x04 / BA + 0x08)
Set voltage V
set
from 0 to V
max
in V. If V
set
greater then V
max
(BA + 0x24 / BA + 0x28), V
set
will be not changed.
Ramp speed Channel A/B
(BA + 0x0C / BA + 0x10)
Voltage ramp speed from 2 V/s to 255 V/s. All processor controlled changes in the output voltage are performed
at this ramp speed.
Actual voltage Channel A/B (BA + 0x14 / BA + 0x18)
Output voltage V
out’
of the channels in V.
Actual current Channel A/B (BA + 0x1C / BA + 0x20)
Output current I
out
of the channels corresponding current resolution.
Hardware limits
(BA + 0x24 / BA + 0x28)
D0 .. D3
Maximal output current (I
max
) in 10 %, hardware setting on the front panel switches
D4 .. D7
Maximal output voltage (V
max
) in 10 %, hardware setting on the front panel switches
D8 .. D15
0
Data ready
(BA + 0x2C)
D7 D6 D5 D4 D3
D2
D1
D0
0
0
0
0
Current B
Voltage B
Current A
Voltage A
The individual bits are set as soon as actual measured data is existing. The bits are reset after the
corresponding reading command.