WARNING!
WARNING!
The protective conductor connection must be ensured by an appropriate mains cable. Before connecting to the
local power supply, check whether the nominal voltage of the devices corresponds to the mains voltage.
WARNING!
WARNING!
The mains connection is made with basic insulation and protective conductor. The device may only be operated
with the protective earth conductor (PE) connected!
The protective conductor connections must be checked for proper function after installation.
WARNING!
WARNING!
The high-voltage generation may only be switched on with this device if the corresponding counterpart with
arrester is contacted at the output of the high-voltage supply.
CAUTION!
CAUTION!
For devices equipped with a LEMO high voltage connector (L16, L30), the high voltage must not be switched on
until the corresponding counterpart has been contacted.
CAUTION!
CAUTION!
When installing the units, make sure that an air flow through the corresponding air inlet and outlet openings is
possible.
CAUTION!
CAUTION!
Before connecting or disconnecting the interface cable, the device must be switched off or disconnected from
the mains.
INFORMATION
INFORMATION
Please check the compatibility with the devices used.
INFORMATION
INFORMATION
It is strongly recommended to read the manual before operation!
THQ Systems - CONFIGURABLE AC/DC LABORATORY HV SUPPLY | Last modified on: 2022-04-21 | www.iseg-hv.com
6/41