9.5 Input Examples
Commands
Explanations
query
#1
Read identification
answer
600138;2.01;3000;405
serial number 600138, firmware version 2.01,
nominal voltage 3000 V, nominal current 40
∙
10
5
nA =
4 mA
Explanations: 405 = 40
∙
10
5
∙
10
-9
= 40
∙
10
-4
= 4
∙
10
-3
= 4mA
command
D1=1000
Set voltage on channel 1 to 1000 V (device goes into USB mode)
command
C1=1E-3
Set current limitation on channel 1 to 1 mA
query
U2
Read measured voltage on channel 2
answer
999.7
999.7 V
query
I1
Read measured current on channel 1
answer
0.028E-3
0.028E-3 = 0.028
∙
10
-3
= 28 µA
query
S3
Read status of channel 3
answer
31
HV ON + NE USB
9.6 Compatibility with THQ 1.xx
The instruction set was simplified with Version 2.00. It is recommended to use the new instruction set for communication with
the device. However, it is possible to emulate the behavior of THQ firmware 1.xx. To do so, the command E1=2 has to be sent to
the device. If the device has more than one channel, the command has to be executed for every channel.
The following differences apply in compatibility mode:
•
For every answer, the command is repeatet first
•
The current limit is transmitted in mA (for I
nom
≥ 1 mA) respectively in µA (for I
nom
< 1 mA).
query
#1
Query identifier
answer
#1
answer
600123;2.01;5000;205
serial 600123, Version 2.01, 5000 V, 2 mA
query
C1=2
Set current limit in channel 1 to 2 mA
answer
C1=2
query
C1
Query current limit in channel 1
answer
C1
answer
2.0
Current limit = 2 mA
Table 7: Examples
THQ Systems - CONFIGURABLE AC/DC LABORATORY HV SUPPLY | Last modified on: 2022-04-21 | www.iseg-hv.com