Event pacing uses the combination of both the programmed enable
conditions and the external enable (pin F of the external flow meter
connector) to determine the enable state. Each time it becomes en-
abled, the sampler takes one sample, placing it in one bottle. The
sampler must become disabled between events. The sampler always
takes a sample at the start time for event paced programs. Nonuni-
form clock time programming schedules each sample event individ-
ually. The sampler skips samples scheduled while it is disabled.
Although you can combine flow pacing and all time pacing types
with any distribution, event paced programs by definition uses only
sequential distribution. The sampler finishes an event paced pro-
gram after depositing a sample in each bottle.
Distribution
Distribution describes how the sampler is to deposit samples. A
sample is the volume of liquid deposited in a bottle. A sample event
includes the full sampling cycle and may deposit a sample into
more than one bottle. You can program the sampler for five distri-
bution methods:
• Sequential
• Composite
• Bottles per sample
• Multiple Bottle Compositing
• Samples per bottle
Sequential
In sequential distribution, the sampler deposits one sample in each
bottle. A sequential sample represents a “snapshot” of the flow
stream at a point in time.
Bottles Per Sample
In bottles per sample distribution, the sampler deposits a sample
in each of a set of bottles. A bottle set includes at least two bottles
but may include all bottles. Use bottles per sample when the vol-
ume to be collected is larger than the amount one bottle can hold or
when you need identical samples.
Samples Per Bottle
In samples per bottle distribution, the sampler deposits samples
from several sample events in a single bottle before moving to the
next bottle. Use samples per bottle distribution to collect a series of
small composite samples.
Composite
For single bottle configurations, samples per bottle distribution is
known as composite sampling. A composite sample represents an av-
erage of the flow stream’s characteristics during the sampling period.
Distribution
Menu Charts appear in:
Figure 23, Standard Programming:
Summary of Contents for 6700SR
Page 2: ......
Page 4: ......
Page 18: ...6700SR Refrigerated Sampler iv...
Page 148: ...6700SR Refrigerated Sampler 130 Appendix B Material Safety Data Sheets...
Page 149: ...6700SR Refrigerated Sampler 131...
Page 150: ...6700SR Refrigerated Sampler 132...
Page 151: ...6700SR Refrigerated Sampler 133...
Page 152: ...6700SR Refrigerated Sampler 134...
Page 153: ...6700SR Refrigerated Sampler 135...
Page 154: ...6700SR Refrigerated Sampler 136...
Page 155: ...6700SR Refrigerated Sampler 137...
Page 156: ...6700SR Refrigerated Sampler 138...
Page 157: ...6700SR Refrigerated Sampler 139...
Page 158: ...6700SR Refrigerated Sampler 140...
Page 159: ...6700SR Refrigerated Sampler 141...
Page 160: ...6700SR Refrigerated Sampler 142...
Page 161: ...6700SR Refrigerated Sampler 143...
Page 162: ...6700SR Refrigerated Sampler 144...
Page 163: ...6700SR Refrigerated Sampler 145...
Page 164: ...6700SR Refrigerated Sampler 146...
Page 165: ...6700SR Refrigerated Sampler 147...
Page 166: ...6700SR Refrigerated Sampler 148...
Page 167: ...6700SR Refrigerated Sampler 149...
Page 168: ...6700SR Refrigerated Sampler 150...
Page 174: ...6700SR Refrigerated Sampler 156 Appendix D Replacement Parts REFRIGERATOR ASSEMBLY FRONT BACK...
Page 175: ...6700SR Refrigerated Sampler 157 REFRIGERATOR ASSEMBLY...
Page 176: ...6700SR Refrigerated Sampler 158 MODEL 6700 CONTROLLER INSIDE...
Page 177: ...6700SR Refrigerated Sampler 159 MODEL 6700 CONTROLLER INSIDE...
Page 178: ...6700SR Refrigerated Sampler 160 MODEL 6700 CONTROLLER OUTSIDE...
Page 179: ...6700SR Refrigerated Sampler 161 MODEL 6700 CONTROLLER OUTSIDE...
Page 180: ...6700SR Refrigerated Sampler 162 MODEL 6700 PUMP GEAR CASE ASSEMBLY...
Page 188: ...6700SR Refrigerated Sampler 170...
Page 190: ......