![ISC GasBadge Plus 3.0 Manual Download Page 36](http://html1.mh-extra.com/html/isc/gasbadge-plus-3-0/gasbadge-plus-3-0_manual_2092319036.webp)
36
INDUSTRIAL SCIENTIFIC
Revision 5 (p/n: 17120932)
displayed. The entire display blinks. In addition, the instrument will be in a high
alarm condition during this state.
From this screen, holding both buttons for 10 seconds allows you to power down
the instrument permanently in preparation for returning it to Industrial Scientific
for repairs.
WARNING:
If this error occurs, you must shut down the personal monitor and
return it to Industrial Scientific for servicing. Once the personal monitor is shut
down, it cannot be reactivated, until it has been serviced by Industrial Scientific.
Unexpected Instrument Error
The Unexpected Instrument Error screen
indicates that an unexpected instrument error
has occurred and the instrument will reset. If
the error persists, the instrument must be
returned to Industrial Scientific for repair or
replacement.
The instrument is in periodic alarm (one alarm
burst every 15 seconds) while in this condition,
to attempt to conserve battery life.
Unexpected Error Screen
Holding both buttons for 10 seconds from this screen allows you to power the
instrument down to return it to Industrial Scientific.
WARNING:
If this error occurs, you must shut down the personal monitor and
return it to Industrial Scientific for servicing. Once the personal monitor is shut
down, it cannot be reactivated, until it has been serviced by Industrial Scientific.