Revision 5 (p/n: 17120932)
INDUSTRIAL SCIENTIFIC
21
Days Since Calibration
The Days Since Calibration screen
indicates the number of days that have
passed since the last calibration. The Days
Indicator is active to designate this number
as days and the Gas Bottle indicator is
active to designate it as a calibration issue.
Pressing the Mode button at this screen will
navigate to the Initiate Print screen.
Days Since Calibration Screen
Actions Available from Days Since Calibration Screen
Action Response
Mode button press
Go to Initiate Print
Enter button press
No response
Mode and Enter buttons
held (3 sec.)
+
Go to Gas Monitoring screen
30 second timeout
Go to Gas Monitoring screen
Event Log
The GasBadge Plus stores alarm events in non-volatile memory. The last 15 gas
alarm events are stored with continuous-loop logging.
The information stored for
each event is:
gas type
cumulative alarm time prior to alarm event
peak exposure level (ppm or %)
battery life remaining prior to alarm event
alarm duration in minutes/seconds
relative time that alarm occurred.
Continuous-loop logging means that once the maximum number of alarm events
has been recorded, the instrument automatically overwrites the oldest alarm event.
The event log can be printed directly to an infrared point-of-sale (POS) printer
available from Industrial Scientific, or it can be downloaded to a PC via a Datalink
or Cal Plus™ accessory product.