The STLG option characteristics are the following:
•
Input: form STS 5000, via the booster connector
•
Output current ranges: 11, 22, 35, 55, 105 A AC. Corresponding no-load voltages: 540, 270, 160, 108, 55 V
•
Output current connection: via high current sockets
•
Output power: 1,800 VA steady; 5,500 VA peak for 5 s, 5,200 VA on the 105 A range
•
High current selector switch, to connect the selected tap to the output current sockets
•
Analogue voltage meter, to display the voltage at the output sockets. Range: 600 V AC
•
Outputs to STS:
•
Selected current output range, to the 10 V voltage input
•
Output current, to the 10 A current input
•
Output voltage, to the 300 V voltage input
•
Transport case: black plastic case, with handles
•
Weight: 25 kg
•
Dimensions: 23 x 33 x 44 cm
All the followings connection cables are included in the option:
•
No. 1 Cable, 6 m long, for the connection of the STS BOOSTER connector
•
No. 2 Cables, 6 m long, for the current output connection, terminated with “Kelvin” type clamps
•
No. 6 cables, red and black, 6 m long
•
No. 1 cable for the connection of the STLG voltage meter output to the STS 300 V meter
•
No. 1 bridge, to short circuit the current meter output if it is not connected to STS
•
Cable transport case
ATTENTION: For safety reasons, STLG must always acts together with the STSG option
4.13
Power factor correction module (code PII85175)
This option is applicable only the STLG option. It is a module to increase the current in the grounding and step and touch
tests; the module has many capacitors and it allows to reduce the reactive component of the line, to obtain the
increasing of the STLG test generated current.
•
Dimensions: 325L, 295A, 285P. Aluminum case with handles
•
Weight: 12 kg
•
Selectable capacitance nominal values: 600 μF, 400 μF, 200 μF, 150 μF and 100 μF
•
Maximum voltage: 600 V (range 100 μF and 150 μF) , 450 V on the others
•
Maximum current: 60 A
Shop for Power Metering products online at:
1.877.766.5412
www.
PowerMeterStore