EMISSIVITY (
ε
)
Range .10 to .99
Two-digit “press to change” digital switch used to adjust the
electronic circuits to match the emissivity to the material being
measured.
Response Time
Range 0.15 to 10 seconds
Potentiometer used to adjust the electronics circuits for the
most suitable response to temperature changes. Use low
setting to "track" rapid changes. Use higher setting to get a
smooth, average temperature display.
EMISSIVITY (
ε
)
Range .10 to .99
Two-digit “press to change” digital switch used to adjust the
electronics circuits to match the emissivity to the material
being measured.
PEAK PICKER PLUG (See Text)
PEAK PICKER DECAY TIME
Range 15 seconds to 30 minutes for Full Scale
Decay
Potentiometer used to adjust the time it takes for a full scale
signal to decay to zero scale value. Rotate in direction of
arrow (clockwise) to increase decay time. Rotate in opposite
direction (counterclockwise) to decrease decay time.
RESPONSE TIME
Range 0.15 to 10 seconds
Potentiometer used to adjust the electronic circuits for the
most suitable response to temperature changes. Use low
setting to "track" rapid changes. Use higher setting to get a
smooth, average temperature display. Rotate in direction of
arrow (clockwise) to increase response time. Rotate in
opposite direction (counterclockwise) to decrease response
time.
Section 3 — OPERATION
3–3
BASIC MODLINE 4
(Model No. Block D = 0)
FIG. 3.1 – QUICK GUIDE TO CONTROLS
MODLINE 4 WITH
PEAK PICKER OPTION
(Model No. Block D = 1)