1–8
Section 1 — INTRODUCTION
The following specifications differ for Series 22 as compared to Series 43 through 48:
Series 43 through 48
Series 22
Calibration Accuracy:
Within 0.75% of full-scale
Within 1.0% of full-scale
temperature or 3
°
F (1.7
°
C),
temperature or 5
°
F (2.8
°
C),
whichever is greater
For the 43-02C and 43-04F:
within 0.75% of full-scale
temperature or 4°F (2.3°C)
whichever is greater
whichever is greater
Repeatability:
0.3% of full-scale temperature
0.5% of full-scale temperature
Temperature Coefficient:
≤
0.015%* of span per
°
F change
≤
0.020% of span per
°
F change
from nominal value of 77
°
F
from nominal value of 77
°
F
over range of 32 to 150
°
F
over range of 32 to 130
°
F
≤
0.027%* of span per
°
C change
≤
0.035% of span per
°
C change
from nominal value of 25
°
C
from nominal value of 25
°
C
over range of 0 to 66
°
C
over range of 0 to 54
°
C
Case Operating
Standard enclosure:
Standard enclosure:
Temperature Range:
32 to 150
°
F (0 to 66
°
C)
32 to 130
°
F (0 to 54
°
C)
(see previous page for exceptions)
Water-cooled enclosure:
Water-cooled enclosure:
32 to 400
°
F (0 to 204
°
C)
32 to 400
°
F (0 to 204
°
C)
Environmental Rating:
NEMA 4
NEMA 4
Optical Centering:
Optical axis within 1
°
of
Optical axis within 2
°
of
mechanical center axis (within
mechanical center axis (within
0.2 in. per foot or 17 mm per
0.4 in. per foot or 33 mm per
meter separation distance)
meter separation distance)
*For Series 43-04F:
≤
0.05% per
°
F
*
For Series 43-02C:
≤
0.09% per
°
C