14
T1R - Recording Sensor
Operation
Connecting External Power
The T1R comes with two AA internal batteries. This provides enough power for at least
one year of operation at the rate of four measurements per hour.
If auxiliary power is desired, you can use a 6 - 13 VDC supply that can provide 15
mA. Connect to Vaux++ (pin 1 - white) and Ground (pin 5 - blue) or contact INW for
auxiliary power supplies.
Connecting the T1R to a Computer
Connect the T1R to your computer’s USB port, as shown below. Drivers and
instructions come with the adapter. (For alternate connection options, see Alternate
Connection Options Section.)
Connect the sensor to your PC using INW’s USB to RS485 adapter.
(See Appendix C for alternate connections.)
PC or Laptop
Computer
USB Port
USB to RS485
Adapter
Sensor
Sensor
Cable