Invasix BodyTite™ Service Manual DO600463F
0344
Page 50 of 52
RF test
After the RF calibration is finished and settings were saved into
system, do not press the OK key
or turn off the system.
Press the footswitch and observe that the
“Power”
display is in range of +/- 15%
of the
“Setup Power”.
Change
“Setup Power”
to 40W energy and press the footswitch’ observe that the
“Power”
display is in range of +/- 10% of the “Setup Power”.
Change
“Setup Power”
to 75W energy and press the footswitch’ observe that the
“Power”
display is in range of +/- 10% of the “Setup Power”.
Press the OK key
to finish the test, a “TEST COMPLETE” screen
appears. Write down all parameters showing on the screen.