Invasix BodyTite™ Service Manual DO600463F
0344
Page 37 of 52
Power Supply Disassembling
Remove front cover as shown in the Front cover removal section
Remove the 20 pin Molex connector
Remove the AC input connector
Remove the GND Quick disconnection
Open the 4 screws attaching the power supply to the cabinet
Remove the power supply from cabinet
Power Supply Assembling
Position the power supply on its standoffs and attach it to the frame using the four
screws
Insert the 20 pin Molex connector to the power supply
Insert the AC power connector into the power supply
Insert the GND quick disconnection to the GND terminal on the power supply located
to the right of the AC power connector
Power Supply
4 x Screws
20 pin Molex connector
AC power connector
GND Quick disconnection