Section 3 - System Installation
Electrical Requirements
The System will require separate line supply of: single phase 90-254 VAC; 5A; 50-
60Hz.
Power receptacles must be within 15 feet of the System site.
The System should not share a power line with other equipment.
For continued protection against fire, replace the fuse only with one of the
same type and rating.
Proper grounding is essential for safe operation.
Environmental Requirements
Corrosive materials can damage electronic parts; therefore, the System should operate
in a non-corrosive atmosphere.
For optimal operation of the System, maintain room temperature between 20º-27ºC
(68º-79ºF) and relative humidity of less than 80%.
Equipment List
The System includes the following:
System platform
TiteFx hand piece
Foot Switch
Operator Manual
Power cord