Invacare Standing Transfer Vest User Manual Download Page 4

Invacare Standing Transfer Vest

WARNING!
Risk of Injury or Damage

Ignition sources can cause burns or fire.

– Patient transfer must be performed with a

safety space between the lift and possible
ignition sources (heater, stove, fireplace, etc.).

– The patient and assistants must not smoke

during transfer.

– The sling must not be placed over heat sources

(heater, stove, fireplace, etc.).

CAUTION!
Compatibility of Slings with Attachment System

Invacare uses a common attachment system
based on hooks and loops. Loops on the sling are
attached to hooks on the lift.

– A risk assessment is always to be carried out

by a professional prior to use of Invacare slings
with lifts of other manufacturers.

– If you are in doubt about the use of Invacare

slings with equipment of other manufacturers,
contact your Invacare provider for advice.

2.2 Labels and Symbols on the Product

The labels are placed on the upper backside of the sling.

Symbols — Fig. 1

A

Read user manual

B

European Conformity

C

Manufacturer

D

Medical device

E

Reference number

F

Serial number

G

Date of manufacture

H

Lot number

I

Max. safe working load

J

Do not bleach

K

Tumble dry, low temperature

L

Do not iron

M

Washing max. 90 °C

N

Field for patient name

O

QR code with link to instruction video

3 Product Overview

3.1 Intended Use

The Invacare Standing Transfer Vest is a non-rigid
body-support unit to transfer an individual from one seating
or standing position to another. The sling is intended to
be combined with a ceiling hoist using a loop to hook
attachment with 2 connection points.

The sling is not intended to be left underneath the patient
after the transfer.

A healthcare professional or private person who has received
proper training is the intended operator of this product.

Indications

The slings are indicated for the transfer of patients with the
ability to stand up, with some degree of weight bearing ability
and with the following minimum degrees of body control:

Leg control / weight
bearing ability

Hip
control

Trunk
control

Head
control

Good

Good

Good

Good

The sling is contraindicated for individuals who are not
cooperative or are not able to understand and follow
instructions.

3.2 Accessories

Extension loops – 20 cm or 30 cm length
Intended to be attached to the outer loops and extend
the suspension straps of the sling for more reach or
positioning options.

Groin Band – S or L
Intended to be attached to the sling and applied around
the patients thighs, in order to keep the sling in place
and prevent it from sliding upwards.

4 Usage

4.1 General Safety Information

WARNING!
Risk of injury or damage

Improper use of this product may cause injury
or damage.

– Do not attempt any transfer without approval

of the patient’s healthcare professional.

– Ensure that the patient is not prone to spasm

or a risk assessment has been completed to
address the possibility of forward pitching.

WARNING!
Risk of injury

Improperly attached or adjusted slings can cause
the patient to fall or cause injury to assistants.

– See the user manual of the lift in addition and

follow the instructions for safety and use.

– Check for proper sling attachment before

starting the transfer.

– Check again for proper sling attachment when

the straps are tensioned before lifting the
patient off the surface.

– Do not use any kind of plastic back incontinence

pad or seating cushion between patient and
sling material that may cause the patient to
slide out of the sling during transfer.

– Make sure that there is sufficient head support

when lifting a patient.

– Position the patient in the sling as directed by

the instructions provided with the sling.

– Adjustments for patient safety and comfort

should be made before moving the patient.

WARNING!
Risk of injury

Usage of damaged slings can cause the patient to
fall or cause injury to assistants.

– Bleached, torn, cut, frayed, or damaged slings

are unsafe and could result in injury. Discard
immediately.

To select the appropriate sling, a risk assessment must be
performed by a healthcare professional. The risk assessment
must consider:

The patient’s weight, size, physical ability and medical
condition.

The type of transfer and the environment.

The compatibility to the other lifting equipment used.

4

60124418-A

Summary of Contents for Standing Transfer Vest

Page 1: ...ransfer Vest en Sling User Manual 3 da Sejl Brugsanvisning 6 de Patientengurt Gebrauchsanweisung 9 es Eslinga Manual del usuario 12 fi Nostoliina K ytt ohje 16 fr Sangle Manuel d utilisation 19 it Imb...

Page 2: ...Fig 1 Fig 4 A B C D E F G H I J K L M XXXXXXXX ISO 10535 XS S M L XL Invacare Portugal Lda Rua Estrada V elha 949 4465 784 Le a do Balio Portugal N O XXXXXXXX XXXXXXXX HMI XXXXX XXXkg XXXXX X YYYY MM...

Page 3: ...er manual Incorrect use Natural wear and tear Incorrect assembly or set up by the purchaser or a third party Technical modifications Unauthorised modifications and or use of unsuitable spare parts 1 5...

Page 4: ...ead control Good Good Good Good The sling is contraindicated for individuals who are not cooperative or are not able to understand and follow instructions 3 2 Accessories Extension loops 20 cm or 30 c...

Page 5: ...d be subject to safety inspection on a six monthly basis Please refer to the HSE web site for guidance www hse gov uk The person responsible for the equipment must ensure adherence to LOLER regulation...

Page 6: ...shyppigheden vaskemetoder og den v gt der b res 1 3 Garantioplysninger Vi yder en producentgaranti p produktet i overensstemmelse med vores almindelige forretningsbetingelser og vilk r i de respektive...

Page 7: ...illingsdato H Partinummer I Maks sikker arbejdsbelastning J M ikke bleges K T rretumbling lav temperatur L M ikke stryges M Vask max 90 C N Felt til patientnavn O QR kode med link til instruktionsvide...

Page 8: ...stg relse af sejlet p personl fteren Sejlets suspensionsstropper er udstyret med farvekodede l kker der giver forskellige l ngder Afstem l kkernes farver efter de tilsvarende stropper p hver side af s...

Page 9: ...dem Produkt informieren Sie bitte den Hersteller und die zust ndige Beh rde in Ihrem Land Symbole in diesem Dokument In diesem Dokument werden Symbole und Signalw rter verwendet um auf Gefahren oder u...

Page 10: ...lifter und m glichen Z ndquellen Heizung Ofen Kamin usw bestehen Weder Patient noch Begleitpersonen d rfen w hrend des Transfers rauchen Der Patientengurt darf nicht ber Z ndquellen Heizung Herd Ofen...

Page 11: ...ARNUNG Verletzungsgefahr Die Verwendung besch digter Patientengurte kann dazu f hren dass der Patient herunterf llt oder Begleitpersonen verletzt werden Ausgeblichene verschlissene eingeschnittene aus...

Page 12: ...Detaillierte Informationen siehe 5 Instandhaltung Seite 11 Stellen Sie sicher dass die Gebrauchsanweisung zusammen mit dem Produkt bergeben wird Das Produkt darf nicht wiederholt verwendet werden wen...

Page 13: ...a pruebas y cumple con la norma ISO 10535 Gr as para el traslado de personas con discapacidad y todos los est ndares relacionados Para obtener m s informaci n sobre las normativas locales p ngase en...

Page 14: ...sostener el peso Control de la cadera Control del tronco Control de la cabeza Bueno Bueno Bueno Bueno La eslinga est contraindicada para personas que no sean capaces de colaborar o de comprender y seg...

Page 15: ...e si alguna pieza muestra signos de deterioro desgaste o un posible fallo Inspecci n peri dica IMPORTANTE Las inspecciones las debe realizar una persona cualificada que conozca bien el dise o el uso y...

Page 16: ...st VAROITUS Osoittaa vaarallisen tilanteen joka voi johtaa vakavaan vammaan tai kuolemaan jos sit ei v ltet HUOMIO Osoittaa vaarallisen tilanteen joka voi johtaa pieneen tai liev n vammaan jos sit ei...

Page 17: ...rviointi aina ennen kuin Invacare nostoliinoja k ytet n muiden valmistajien potilasnostinten kanssa Invacaren nostoliinojen k ytt misest muiden valmistajien laitteiden kanssa saat lis tietoa Invacare...

Page 18: ...inaa alas potilaan sel n taakse keskelle kunnes sen yl reuna on potilaan lapaluiden alapuolella Fig 2 3 Nojaa potilasta taakse nostoliinaa vasten 4 Sulje vy potilaan rintakeh n ymp rill ja varmista et...

Page 19: ...le des caract res de la version imprim e du document vous semble trop difficile lire vous pouvez t l charger la version PDF sur le site Internet Vous pourrez alors ajuster la taille des caract res l c...

Page 20: ...de levage Consultez la documentation ou l tiquetage pour conna tre la charge maximale d utilisation Le composant ayant la limite de charge la plus basse d termine la charge maximale d utilisation de l...

Page 21: ...nt fix e une fois les bandes tendues avant de soulever le patient de la surface sur laquelle il se trouve N utilisez aucun type de serviette d incontinence plastifi e ni coussin d assise entre le pati...

Page 22: ...instructions de lavage figurant sur l tiquette L utilisation fr quente de temp ratures lev es et le s chage en machine peuvent r duire la dur e de vie du produit Nous recommandons l utilisation d une...

Page 23: ...he tecniche Modifiche non autorizzate e o utilizzo di pezzi di ricambio non adatti 1 5 Conformit La qualit fondamentale per l azienda che opera nel rispetto e nell ambito della norma ISO 13485 Il prod...

Page 24: ...nit di supporto del corpo non rigida per trasferire un individuo da una posizione seduta o in piedi ad un altra L imbracatura destinata a essere abbinata a un sollevatore a soffitto mediante un occhie...

Page 25: ...izionamento delle braccia del paziente Le braccia del paziente devono essere posizionate tra i punti di fissaggio delle cinghie di sospensione 4 3 Fissaggio dell imbracatura al sollevatore Le cinghie...

Page 26: ...t leest of u de juiste versie hebt U vindt een pdf bestand met de laatste versie op de Invacare website Als u de lettergrootte in de gedrukte versie van het document moeilijk kunt lezen kunt u een pdf...

Page 27: ...e of op het label Het onderdeel met de laagste maximale belastbaarheid is bepalend voor de maximale veilige belasting van het gehele systeem WAARSCHUWING Risico op schade of lichamelijk letsel Ontstek...

Page 28: ...vlak opnieuw of de draagband goed is bevestigd Gebruik geen enkel type kunststof incontinentieluier of zitkussen tussen de pati nt en het materiaal van de draagband waardoor de pati nt tijdens het ver...

Page 29: ...oge temperatuur of drogen in de droger kan de levensduur van het product worden verkort Wij adviseren een normaal wasmiddel voor huishoudelijke doeleinden te gebruiken Gebruik geen wasverzachter Kies...

Page 30: ...sk utstyr klasse 1 Lanseringsdatoen for dette produktet er oppgitt i CE samsvarserkl ringen Vi arbeider kontinuerlig for sikre at selskapets p virkning p milj et b de lokalt og globalt reduseres til e...

Page 31: ...v re festet til de ytre l kkene og til forlenge fj ringsstroppene p slyngen for lenger rekkevidde eller posisjoneringsalternativer Lyskeb nd S eller L Ment v re festet til l fteseilet og p f res rundt...

Page 32: ...enter skades Inspiser l fteseilet for slitasje og l s s m etter hver vask VIKTIG Regelmessig rengj ring og desinfisering forhindrer forurensning Rengj r og desinfiser produktet jevnlig mens det er i b...

Page 33: ...gaste natural devido ao uso Montagem ou prepara o incorreta pelo comprador ou por terceiros Modifica es t cnicas Modifica es n o autorizadas e ou utiliza o de pe as sobressalentes desadequadas 1 5 Con...

Page 34: ...anding Transfer Vest uma unidade de suporte corporal n o r gida destinada transfer ncia de um indiv duo de uma posi o sentada ou de p para outra A cesta destina se a ser combinada com um guincho de te...

Page 35: ...a os do paciente Os bra os do paciente devem ser colocados entre os pontos de fixa o das correias de suspens o 4 3 Fixa o da cesta ao elevador de transfer ncia As correias de suspens o da cesta est o...

Page 36: ...versionen som PDF fil p Invacares webbplats Om du tycker att teckensnittsstorleken i den tryckta versionen av bruksanvisningen r sv r att l sa kan du ladda ned PDF versionen fr n webbplatsen Du kan s...

Page 37: ...st nd mellan personlyften och m jliga ant ndningsk llor element spis eldstad etc Patienten och v rdpersonalen f r inte r ka under f rflyttningen Lyftselen f r inte placeras ver v rmek llor element spi...

Page 38: ...a tillst nd Typ av f rflyttning och omgivningen Kompatibiliteten med vrig lyftutrustning som anv nds Mer information finns i Guide f r val av Invacare lyftsele 4 2 Applicera lyftselen Applicera v sten...

Page 39: ...re Inspektion Reng ring och desinfektion F r ing ende information se 5 Underh ll Sida 38 Se till att bruksanvisningen verl mnas tillsammans med produkten Om skador eller fel uppt cks ska produkten int...

Page 40: ...445 38 00 34 italia invacare com www invacare it Nederland Invacare BV Galvanistraat 14 3 NL 6716 AE Ede Tel 31 0 318 695 757 Fax 31 0 318 695 758 nederland invacare com www invacare nl Belgium Luxemb...

Reviews: