Rea® Focus™
5 mm allen key
Location of the rear wheel:
C
D
E
1.
Loosen the screws
C
.
2.
Turn the rear wheel clamps
E
around to position the
rear wheel behind the rear wheel bar
D
.
3.
Re-tighten the screws
C
.
5 mm allen key
If the rear wheels are positioned in front of the
rear wheel bar
D
, the wheelchair will become more
unstable. If the rear wheels are positioned further
back, the stability will increase.
Camber
Depending on the rear wheel attachment, this
product can have 0 or 2 camber. To achieve 2 camber,
the rear wheel attachment needs to be changed.
E
C
F
1.
Loosen the screws
C
.
2.
Remove the rear wheel attachment clamps
E
.
3.
Mount the camber rear wheel attachment
F
.
The 2 camber rear wheel attachment has a round
marking on the outside of the clamp. Make sure
that this marking is positioned upwards in order
to obtain the correct angle.
4.
Tighten the screws.
5 mm allen key
Adjusting the castors
WARNING!
Risk of tipping
The castor might come off.
– Make sure to check that the castors are securely
locked after each adjustment.
WARNING!
Safety risk
– Make sure that the castor housing is properly
secured after each adjustment.
When you have found the desired seat height and
depth, it’s important to check that the angle between
the castor attachment and surface is as close to
90° as possible, as it may affect the wheelchairs
propelling ability.
Castor angle
1.
A
B
Loosen the screw
A
.
2.
Loosen the screw
B
two turns.
3.
C
Pull out the adjustment screw
C
.
20