SECTION 1—GENERAL GUIDELINES
Platinum
™
Series
6
Part No 1118353
A spontaneous and violent ignition may occur if oil, grease or
greasy substances come in contact with oxygen under
pressure. These substances MUST be kept away from the
oxygen concentrator, tubing and connections, and all other
oxygen equipment. DO NOT use any lubricants unless
recommended by Invacare.
For optimum performance, Invacare recommends that each
concentrator be on and running for a minimum of 30 minutes
at a time. Shorter periods of operation may reduce maximum
product life.
If the concentrator has a damaged cord or plug, if it is not
working properly, if it has been dropped or damaged, or
dropped into water, call qualified technician for examination
and repair.
Keep the cord away from heated or hot surfaces.
DO NOT move or relocate concentrator by pulling on the
cord.
NEVER drop or insert any object into any opening.
NEVER block the air openings of the product or place it on a
soft surface, such as a bed or couch, where the air opening
may be blocked. Keep the openings free from lint, hair and the
like. Fill humidifier with water to the level shown by the
manufacturer. DO NOT overfill.
Keep unit at least three (3) inches away from walls, draperies,
furniture, and the like.
Invacare recommends that Crush-Proof oxygen tubing be
used with this product and not exceed 50 ft. (15.2m) in length.
Recommended use for Platinum 10 is up to 25 ft. in length of
high flow tubing.