SECTION 7—TROUBLESHOOTING GUIDE
Part No 1118353
33
Platinum
™
Series
YELLOW or RED
LIGHT Illumi-
nated.
Alarm:
CONTINUOUS
On red light only.
* Only applies to
5LXO
2
and
10LXO
2
SensO
2
models.
1. Low oxygen purity.*
2. Kinked or blocked tub-
ing, cannula or humidi-
fier.*
3. Flowmeter set at 0.5
L/min. for Platinum 5 and
1 L/min. for Platinum 10
or less.*
4. Unit overheating due
to blocked air intake.
5. Insufficient power at
outlet.
6. Internal repairs
required.
1. Clean or Replace filters.
2. Inspect for kinks or block-
ages. Correct, clean or
replace item. Once cor-
rected, turn power OFF for
60 seconds and then turn
power back ON.
3. Check flowmeter is set to
1.0 L/min. or more. (Refer to
Recommended Guidelines for
Optimum Performance on
page 22.
4.a. Remove and clean cabinet
filters.
4b. Move concentrator at
least three (3) inches from
walls, draperies, and furni-
ture.
5. Do not use extension
cords. Move to another elec-
trical outlet or circuit.
6. Call Service Provider.
GREEN LIGHT
with YELLOW
Light Flashing.
1. Internal repairs
required.
1. Call Service Provider.
SYMPTOM:
PROBABLE CAUSE:
SOLUTION: