SECTION 5—OPERATING INSTRUCTIONS
Part No 1118353
25
Platinum
™
Series
7. Attach the cannula/patient supply tubing to the
humidifier bottle outlet.
8. After assembly, ensure that oxygen is flowing through the
cannula.
FIGURE 5.4
Attaching the Oxygen Tubing
WARNING
For optimum performance, Invacare recommends that
each concentrator be on and running for a minimum of 30
minutes at a time. Shorter periods of operation may
reduce maximum product life.
Power Switch
NOTE: For this procedure, refer to FIGURE 5.5.
1. Press power switch to on
position. All the panel
lights and the audible
alarm will come on for
one second, indicating
that the unit is
functioning properly.
After one second, only
the GREEN system
ok/power light will stay
on.
FIGURE 5.5
Power Switch
Oxygen
Tubing
Humidifier Outlet
Oxygen Outlet
Connector
On/Off (I/O)
Power Switch