Service
The controls system is supplied with a standard drive
program. If you have carried out customer-specific
modifications to the drive program, you will have to
make these changes again after installing the new
power module.
When disassembling, note position of small parts
such as screws and washers. Put small parts down
so that they can be reassembled in right sequence.
The plugs on the electronic module cannot be
wrongly connected because all plugs have a different
size and only fit in one socket.
•
10 mm wrench
•
17 mm wrench
•
Rubber hammer
•
Phillips screwdriver
•
Supporting wooden block
•
To adapt the drive program: programming
software or hand programming device and
electronic system installation manual, available
from Invacare.
1.
Secure mobility device against rolling away.
2.
Place supporting wooden block under frames at battery
holder height.
3.
Remove seat.
4.
Remove shroud.
5.
Disconnect battery cable (1).
6.
Loosen and remove rear wheel fixing screws.
7.
Remove rear wheels.
8.
Remove drive unit. See 4.10.1 Replacing drive unit, page
27.
9.
Remove screws (1) on power module (2).
•
Rhino1:
DS72KB01/DS112KB02
•
Rhino 2:
DS90/DS120:
10.
Replace power module (2).
11. Install parts in reverse order.
12. Load drive program onto power module. See controls
software description.
13. Test all functions (trial run).
4.5.5 Setup settings
Operating console (LED version)
Arrangement
1608032-A
13