5 Settings and Adjustments
5.1 Measuring correctly
When measuring dimensions such as the seat depth, you
need to measure from defined points to ensure that you
get correct values.
Note that measured values may vary up to ± 10
mm.
Measuring seat depth
Measuring seat width
Measuring backrest height
5.2 Setting Drive Lockout and Limit
Switches
Setting the Drive Lockout for Tilt/Recline Systems — ACS2
Systems
All PPS are supplied with a drive lockout (DLO) switch. This
prevents the drive motors from operating if the combined
tilt/recline angle is positioned beyond a pre-determined
angle. The maximum drive lockout angle is 60° (TDX
SP2)/50° (Storm
4
) from the vertical (see diagram below).
On the Storm
4
, an additional switch activates the DLO if
the tilt angle exceeds 25°.
Fig. 5-1
– Ensure base is on a level surface when
adjusting limits.
After the angle has been set, ensure the drive lockout is
tested. Readjustment may be required.
Fig. 5-2
The DLO switch
A
is typically mounted to the side of the
actuator module on the backrest plate. The additional DLO
switch on the Storm
4
B
is typically mounted under the
shroud on the most rearward point of the seat plate.
If the DLO is activated while adjusting the switch
angle, you must activate the tilt or recline once
to be able to re-test the DLO. Otherwise, the DLO
remains activated, regardless of the switch angle.
The DLO limit should always be set to the angle
that best meets the individual needs of the user
and overall stability of the wheelchair. However,
the DLO angle should not exceed the maximum 60°
(TDX SP2)/50° (Storm
4
) backrest angle.
Setting the Drive Lockout for Tilt/Recline Systems — LiNX
Systems
All PPS are supplied with a drive lockout (DLO) switch. This
prevents the drive motors from operating if the combined
tilt/recline angle is positioned beyond a pre-determined
angle. The maximum drive lockout angle is 60° (TDX
SP2)/50° (Storm
4
) from the vertical (see diagram below).
On the Storm
4
, an additional switch activates the DLO if
the tilt angle exceeds 25°.
Fig. 5-3
– Ensure base is on a level surface when
adjusting limits.
The DLO switch is one of several angle sensors installed in
the actuator module. The DLO angle is set via the
Lockout
1586649-F
29
Summary of Contents for Motion Concepts Ultra Low Maxx
Page 15: ...Overview Fig 3 32 A DX2 ACT2 or DX2 ACT4 1586649 F 15 ...
Page 17: ...Overview Wiring diagram lifter tilt recline TDX SP2 1586649 F 17 ...
Page 19: ...Overview Wiring diagram lifter tilt recline and attendant control TDX SP2 1586649 F 19 ...
Page 21: ...Overview Wiring diagram lifter tilt recline Storm4 1586649 F 21 ...
Page 23: ...Overview Wiring diagram lifter tilt recline and attendant control Storm4 1586649 F 23 ...
Page 97: ...Notes ...
Page 98: ...Notes ...
Page 99: ...Notes ...