Invacare microAIR MA90 Series Owner'S Operator And Maintenance Manual Download Page 18

SECTION 2—OPERATION

MA90 Series

18

Part No 1148139

Mode Button (MA90Z)

1. Press

 

the

 

Mode

 

button

 

(

)

 

to

 

select

 

the

 

Max

 

Inflate

 

Mode

 

or

 

the

 

Low

 

Air

 

Loss

 

mode.

• Max

 

Inflate

 

Mode

 ‐ 

In

 

this

 

mode,

 

the

 

mattress

 

inflates

 

rapidly

 

to

 

maximum

 

firmness

 

(pressurized

 

to

 

35

 

±

 

6

 

mmHg).

 

A

 

series

 

of

 

beeps

 

sound

 

every

 

three

 

minutes

 

as

 

a

 

reminder

 

that

 

the

 

Max

 

Inflate

 

mode

 

is

 

active.

 

After

 

30

 

minutes,

 

the

 

Max

 

Inflate

 

mode

 

deactivates

 

and

 

the

 

control

 

unit

 

defaults

 

to

 

the

 

previous

 

setting.

 

NOTE:

 

It

 

takes

 

5

10

 

minutes

 

for

 

the

 

MA90Z

 

mattress

 

to

 

inflate

 

fully

 

(inflation

 

time

 

depends

 

on

 

size

 

of

 

mattress).

NOTE:

 

Max

 

Inflate

 

mode

 

can

 

be

 

manually

 

disengaged

 

by

 

pressing

 

the

 

Max

 

Inflate

 

button.

 

This

 

will

 

deactivate

 

the

 

Max

 

Inflate

 

LED.

NOTE:

 

It

 

is

 

recommended

 

that

 

Max

 

Inflate

 

setting

 

be

 

used

 

during

 

patient

 

ingress/egress,

 

patient

 

wound

 

care,

 

patient

 

turning

 

or

 

patient

 

cleaning.

• Low

 

Air

 

Loss

 

Mode

 ‐ 

In

 

this

 

mode,

 

the

 

mattress

 

goes

 

into

 

on

demand

 

low

 

air

 

loss

 

relief

 

mode.

Max Inflate Button (MA95Z)

1. Press

 

the

 

Max

 

Inflate

 

button

 

(

)

 

to

 

select

 

the

 

max

 

inflate

 

mode

 

and

 

inflate

 

the

 

mattress

 

rapidly

 

to

 

maximum

 

firmness

 

(pressurized

 

to

 

35

 

±

 

6

 

mmHg).

 

A

 

series

 

of

 

beeps

 

sound

 

every

 

three

 

minutes

 

as

 

a

 

reminder

 

that

 

the

 

Max

 

Inflate

 

mode

 

is

 

active.

 

After

 

30

 

minutes,

 

the

 

Max

 

Inflate

 

mode

 

deactivates

 

and

 

the

 

control

 

unit

 

defaults

 

to

 

the

 

previous

 

setting.

 

NOTE:

 

It

 

takes

 

25

60

 

seconds

 

for

 

the

 

MA95Z

 

mattress

 

to

 

inflate

 

fully

 

(inflation

 

time

 

depends

 

on

 

size

 

of

 

mattress).

NOTE:

 

It

 

is

 

recommended

 

that

 

Max

 

Inflate

 

setting

 

be

 

used

 

during

 

patient

 

ingress/egress,

 

patient

 

wound

 

care,

 

patient

 

turning

 

or

 

patient

 

cleaning.

2. Perform

 

one

 

of

 

the

 

following:

• MA90Z

 ‐ 

Press

 

the

 

Max

 

Inflate

 

button

 

again

 

to

 

manually

 

disengage

 

Max

 

Inflate

 

mode.

 

• MA95Z

 ‐ 

Select

 

a

 

different

 

setting

 

to

 

exit

 

Max

 

Inflate

 

mode.

 

NOTE:

 

This

 

will

 

deactivate

 

the

 

Max

 

Inflate

 

LED.

Summary of Contents for microAIR MA90 Series

Page 1: ...BEFORE using this product read this manual and save for future reference For more information regarding Invacare products parts and services please visit www invacare com MA90 Series MA90Z Alternating...

Page 2: ...facturers have not been tested by Invacare and are not recommended for use with Invacare products SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which coul...

Page 3: ...PERATION 10 Installing the Mattress Replacement System 10 Installing the Side Rails 11 Installing the Power Unit 11 Connecting the Hose 12 Connecting the Power Cord 13 Using the Front Panel 14 Power B...

Page 4: ...g the System 24 Changing the Filter 25 Troubleshooting 25 LIMITED WARRANTY 28 REGISTER YOUR PRODUCT The benefits of registering include 1 Safeguarding your investment 2 Ensuring long term maintenance...

Page 5: ...low fuses MODE OF OPERATION Continuous MA90Z MA95Z WEIGHT CAPACITY STANDARD MATTRESS BARIATRIC MATTRESS 350 lbs 1000 lbs PRESSURE ZONE 2 MAXIMUM FLOW 50 15 LPM 1275 100 LPM MAXIMUM PRESSURE 35 5 mmHg...

Page 6: ...L UNIT CONNECTION Quad Flow Four Single Quick Coupling Triple Flow Two and One Single Magnetic Quick Coupling CONTROL UNIT PACKAGING 1 piece box AIR FILTER None Charcoal Air Filter with Fire Retardant...

Page 7: ...ber WARNING Patient entrapment with bed side rails may cause injury or death Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment Follow the manufacturer s instructions Moni...

Page 8: ...are raised Refer to Installing the Mattress Replacement System on page 10 Entrapment May Occur WARNING Patient entrapment with bed side rails may cause injury or death Mattress MUST fit bed frame and...

Page 9: ...etics Oxygen There is a possible fire hazard when used with oxygen administering equipment other than nasal mask or half bed tent type The oxygen tent should NOT extend below mattress support level El...

Page 10: ...oot ends and to the frame at the center of the bed Otherwise damage to the mattress will occur when the head and foot ends are raised NOTE For this procedure refer to FIGURE 2 1 NOTE The powered mattr...

Page 11: ...g this equipment Invacare product manuals are available at www invacare com or your dealer NOTE Refer to the instructions provided with the side rails for the installation procedure Installing the Pow...

Page 12: ...t 3 Perform one of the following MA90Z Perform the following steps Detail A i Squeeze and hold the tabs on the hose connectors ii Insert the hose connectors into the control unit connector iii Push th...

Page 13: ...tripped over 1 Examine the hospital grade power cord supplied with the control unit 2 Perform one of the following If the plug is damaged Call your supplier for a replacement hospital grade cord If t...

Page 14: ...N LED illuminates indicating that the control unit is on If the power comes on after a power outage the control unit will go through its system initializa tion routine for a few seconds and then resum...

Page 15: ...ngles NOTE The rotation angles are and full rotation 45 Rotation Button MA95Z 1 Press the Rotation button to select Static mode or Alternating Rotation mode NOTE When the LED is on Alternating Rotatio...

Page 16: ...tient s weight to height ratio is above average increase the pressure setting by approximately 20 MA90Z models display levels from 0 to 9 and MA95Z models display the mmHg from 6 to 32 Automatic Wirel...

Page 17: ...is lit and L displays in the LCD Display ii Raise the bed to 45 NOTE The control unit will beep and will return to Standby iii Perform one of the following Cancel Press the Turn Angle button to exit F...

Page 18: ...patient wound care patient turning or patient cleaning Low Air Loss Mode In this mode the mattress goes into on demand low air loss relief mode Max Inflate Button MA95Z 1 Press the Max Inflate button...

Page 19: ...utage NOTE During a power outage the MA90Z mattress retains the air as long as the mattress is connected to the control unit Low Pressure LED In the event that the mattress hose disconnects an alarm s...

Page 20: ...patient is lying flat on his her back in the middle of the mattress B Place four fingers between the air cushions directly underneath the sacral region of the patient s body C Ensure that there is 3 t...

Page 21: ...to the owner s manual provided with the wheelchair 3 Press the Max Inflate button to achieve maximum mattress pressure 4 Raise or lower the bed to match the wheelchair height Refer to the owner s man...

Page 22: ...22 Part No 1148139 FIGURE 2 5 Preparing for CPR Procedure Connector Cover Control Unit Connectors MA95Z Hose Connector DETAIL A Control Unit RED CPR Connector MA90Z Only Mattress DETAIL B DETAIL D MA...

Page 23: ...ganisms 1 Remove the bedding 2 If necessary inflate the mattress 3 Ensure that the control unit is off 4 Unplug the power cord from the wall outlet 5 Ensure that the underside of the mattress is clear...

Page 24: ...ings with a quaternary disinfectant solution B Using a nylon brush gently clean all crevices as they can harbor microorganisms C Air dry all treated surfaces 10 Perform the following steps to clean th...

Page 25: ...trol unit using the two screws FIGURE 3 1 Changing the Filter Troubleshooting PROBLEM CAUSE SOLUTION Mattress not inflating Not alternating properly Mattress hose disconnected Connect hose connectors...

Page 26: ...ck power source and turn on unit Power cord disconnected Connect power cord to the power source No power in the power source Check power source has power power and turn it On Power outage Wait till th...

Page 27: ...SECTION 3 MAINTENANCE AND TROUBLESHOOTING Part No 1148139 27 MA90 Series NOTES...

Page 28: ...ion will issue a return authorization The defective unit or parts must be returned for warranty inspection using the serial number when applicable as identification within thirty days of return author...

Reviews: