SECTION 6—FRONT RIGGINGS
3G Storm Series® Wheelchairs
68
Part No 1134791
Installing/Adjusting Adjustable Angle Flip-Up Footplates
Installing Adjustable Angle Flip-up Footplates
NOTE: For this procedure, refer to FIGURE 6.9.
1. Slide the half clamp over the footplate hinge.
2. Loosely tighten the two flat screws that secure the footplate to the half clamp.
3. Adjust the footplates to the necessary angle and depth for the user. Refer to Installing/
Adjusting Adjustable Angle Flip-Up Footplates on page 68.
Adjusting Adjustable Angle Flip-up Footplates
Depth
NOTE: For this procedure, refer to FIGURE 6.9.
1. Remove the two flat screws, washers
and locknuts that secure articulating
footplate to the half clamp.
NOTE: Observe the angle of the articulating
footplate for reinstallation.
2. Move articulating footplate to one of
four mounting positions.
NOTE: If desired depth is still not obtained,
rotate the half clamp on the footplate hinge 180
o
.
3. Retighten the two flat screws, washers
and locknuts.
NOTE: The settings for positioning the
articulating footplates on the half-clamps may
vary for each footplate.
FIGURE 6.9
Adjusting Adjustable Angle
Flip-up Footplates - Depth
4. Refer to Angle on page 69 or Perpendicular and/or Inversion/Eversion Adjustment on
page 69.
Washers
Locknuts
Half Clamp
Footplate Hinge
Nylon
Adjustment
Screw
Footplate Hinge
Articulating
Footplate
Flat Screws
90° Footrest Support