SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING
Part No 1134791
45
3G Storm Series® Wheelchairs
Wheelchairs with Powered Seating Systems
Joystick erratic or does not respond
as desired.
Damaged motor coupling.
Electrical malfunction.
Controller programmed
improperly.
Contact Dealer/Invacare for Service.
Contact Dealer/Invacare for Service.
Reprogram controller. (Refer to
electronics manual supplied with wheel-
chair).
Wheelchair does not respond to
commands.
Power indicator OFF—even after
recharging.
Poor battery terminal
connection.
Electrical malfunction.
Clean terminals (Refer to Cleaning Bat-
tery Terminals on page 101).
Contact Dealer/Invacare for Service.
SYMPTOM
PROBABLE CAUSE
SOLUTIONS
Wheelchair Power ON
but does not drive
System tilted, reclined or
elevated beyond drive
lock-out angle (20°).
Return to neutral position (upright and completely
lowered).
Refer to Operating Powered Seating
Systems on page 50
.
Contact Invacare/Dealer for service if this does not solve
the problem.
Seating system not
functioning or working
intermittently.
Low batteries.
Faulty electrical
connection.
Blown fuse.
Seat has been driven
under a heavy load for an
extended period of time.
Charge batteries.
Check all connections.
Have wiring harness replaced by a qualified technician.
Allow time for the electronics to cool down (Light Duty
Use).
Programmer does not
work or gives
“communication error”
System tilted, reclined or
elevated beyond drive
lock-out angle (20°).
Return to neutral position (upright and completely
lowered).
Refer to Operating Powered Seating
Systems on page 50
.
Contact Invacare/Dealer for service if this does not solve
the problem.
Wheelchair drives at
reduced speed.
Seating system is
elevated.
Return joystick to neutral position and completely lower
the seat.
SYMPTOM
PROBABLE CAUSE
SOLUTIONS