UV lamp operation time can be set to 5 seconds from 5 seconds to 99 minutes, 55
seconds.
3
Use the top and bottom button to change the time setting.
Opening the door cancels UV disinfection and initialize the procedure. Minimum 5 minutes of UV
disinfection is recommended.
Figure 19. UV setting screen
Warring
Risk of personal injury
UV light is harmful for human. Use carefully.
6. Preheating
Heating Tray starts warming up automatically upon activation of the device (Power on). If you start an
operation, such as a photograph, LCD screen gives a warm-up period with warning signs and you can
begin driving after all this time has elapsed.
Turn on the device 10 minutes before operation to preheat heating blocks. Turning on the device
automatically preheats heating blocks.
1
If preheating system is unfunctional error code appears on LCD screen.
2
Preheated heating blocks is 80
℃
maximum. Do not touch heating block when device is
activated and for 15 minutes after deactivation.
Warring
Risk of personal injury
If heating tray ends warm-up, it has very hot surface. Use carefully.
7. Test mode
'Test mode' is to confirm proper activation of the device before regular operation. It is recommended to
use 'Test Mode' after transportation, certain period of un-used time etc. Installation of strip and plate is
Summary of Contents for Miracle-AutoXT
Page 3: ...Instruction for Use USER MANUAL IBT QMS INT50104 R00 3 34 Section 0 Introduction...
Page 7: ...Instruction for Use USER MANUAL IBT QMS INT50104 R00 7 34 Section II Product Components...
Page 9: ...Instruction for Use USER MANUAL IBT QMS INT50104 R00 9 34 Section III Precautions...
Page 13: ...Instruction for Use USER MANUAL IBT QMS INT50104 R00 13 34 Section IV Installation...
Page 27: ...Instruction for Use USER MANUAL IBT QMS TF 001 R00 27 34 Section VI Troubleshooting...
Page 31: ...Instruction for Use USER MANUAL IBT QMS TF 001 R00 31 34 Section VII Warranty...
Page 33: ...Instruction for Use USER MANUAL IBT QMS TF 001 R00 33 34 Section VIII Symbols...