Heater Wire Connection
Icon
(Heater wire abnormal:
Flashes
red)
Yes
1) Heater Wire is not connected.
2) Heater Wire has come-off or
short-circuited.
1) Connect the heater wire firmly.
2) Replace the heater wire.
Temp Probe Connection
Icon
(Fault: Flashes
red)
Yes
1) Dual Temp Probe is not
connected.
2) Dual Temp Probe has
disconnected or short-circuited
1) Connect the dual temp probe firmly.
2) Replace the dual temp probe.
Icon
(Unit Fault: Flashes
red)
Yes
1) Chamber Temp doesn’t rise more
than 2°C for 30 minutes after
power-on.
2) Heater plate fault
3) Software memory error.
Remove the device from the breathing
system, and contact your sales provider
for repair.
All LED suddenly turn off.
No
Over heat protector is activated.
Remove the device from the breathing
circuit, and contact your sales provider
for repair.
The device is not
activated even though the
alarm condition is
restored.
Yes
Alarm state inter-lock is activated (It
is locked when alarm state
continues for > 10 minutes)
Turn off the device and restart it.
The device is not
powered on.
No
1) Device fused.
2) Heater base fault.
3) Overheat protector is activated.
1) Change the fuse.
2) Remove the device from the
breathing circuit, and contact your
sales provider for repair.
Temp panel is blank or
display
ing “8.8.8.” or
other.
(Display is different
according to the state
when the power is on)
Yes
1)
Failure of self-check
Remove the device from the breathing
circuit, and contact your sales provider
for repair.
* If an alarm state is left unresolved for 10 minutes, then the device displays
“
” and stops operating. The alarm
continues to sound intermittently. The Humidifier base requires turning off and on again to resume operation.
*The priority of each alarm condition is represented by the colour of the LED. Yellow indicates medium priority, RED is high
priority.
14. Compliant Accessories
The following accessories are available to order.
Part Number
Accessory Name
5600000
Electrical adaptor lead (Single heated wire system)
5601000
Electrical adaptor lead (Dual heated wire system)
7700000
Dual temperature probe assembly