Mode Change: 

“IPPV” mode is set as a default when the unit is delivered out from factory.  Keep pressing the “MODE” 
button for 2 seconds, it 

changes with following order:  IPPV→NPPV→MANUAL, so select necessary mode. 

Mode Display LED

:  IPPV→NPPV→MANUAL 

IPPV Mode:

 This mode is for patients with a tracheal tube. The preset for the patient side is 40

C and 

chamber side is 37

C automatically.  

NPPV Mode

: This mode is for patients with NIV (Non Invasive Ventilation) masks. The preset for the patient 

side is 34

C and chamber side is 31

C automatically. 

MANUAL mode

: This mode allows the selection of optional temperature of both patient side and chamber 

side.  (Refer to 5 for the details.) 
On selecting MANUAL mode, a sound confirmation will be activated and the display will show the set 
patient temperature and then the currently set chamber side temperature for approximately 2 seconds each 
as confirmation of the initial setting.  After 2 seconds, LED

 status indication, blinks in green, and the 

patient side temperature can be set.  Temperature setting can be changed by 

 scroll button. While 

pressing 

 

“Set” button, operate 

 scroll button to set the chamber side temperature. 

If you do not make changes for 10 seconds, the unit will return to the normal state, and display will be as 
selected.   
 

Patient side: 30

°

C to 40

°

C (increments 1

°

C) 

Chamber side: 26

°

C to 43

°

C (increments 1

°

C) 

 

Temperature setting of patient side vs. chamber side temperature is in relationship within the range of -

4

°

C to +3

°

C. 

 

The pre-set temperatures are 37

C at chamber side and 37

°

C at patient side when the unit is shipped 

out from the factory.  Refer to 

 section for how to change pre-set value. 

Scroll Button:

  

▲up  ▼down 

In

 

“MANUAL” mode, these buttons allow the user to change the temperature setting of both patient side and 

chamber side. 
Press either ▲up or ▼down button for more than 2 seconds, patient icon will blink quickly in green and 
display patient temperature setting. At this time, temperature setting changes can be made by pressing 
▲up and ▼down button. Press the button once, and the temperature value changes 1

°

C.  Holding the 

button will not change the temperature value continuously. 

Warning ! 
Rain out in the patient circuit will increase when room temperature is lower than patient side set 
temperature. 


 
 
 

Status Indication LED: 

There are 5 LED indicators showing position around the heater base and breathing system on a diagram on 
the front panel.  

1.

 In normal operation the temperature shown in main display (2) is measured from the point indicated by 

LED

 

 

(patient temperature) or LED 

 

(chamber temperature) whichever is illuminated in green. 

2.

  On entering manual mode the temperature shown in display (2) will be the patient temperature when 

LED  

 is blinking green and chamber temperature when LED 

 

is blinking.  To change temp see section 

3.

 Illuminated LED shows location of the fault, as detailed below. 

 

LED

 

 Patient Temperature 

Flashes red

   - the measured temperature has exceeded the set temperature by 2

C, or exceeded a 

maximum temperature of 43

C (in this case there is a high

 

priority audible alarm and the heated wire is 

turned off). 
 

Flashes red

 -  Patient Probe temperature has not increased by >2

C in 2 minutes after Power-on 

(in this case there is a high priority audible alarm).

 

 

Flashes yellow 

– 

the measured temperature is 4

C below the set temperature (in this case a medium 

priority

 

audible alarm sounds intermittently). 

Note:

 this alarm is disabled while the Humidifier base warms up 

 
 
 

LED 

 Chamber temperature 

Flashes

 

red

  - The measured temperature exceeds the set temperature by 4

C for more than 20 minutes, 

or the measured temperature exceeds the set level by 10

C (in this case there is a high priority audible 

alarm and the heated wire is turned off). 
 

Flashes yellow

 

– the measured temperature is 4

C below the set temperature for more than 20 minutes, or 

the measured temperature is below the set temperature by 10

C (in this case a medium priority audible 

alarm sounds intermittently). 

Note:

 this alarm is disabled while the Humidifier base warms up. 

 

Summary of Contents for PMH7000

Page 1: ...03 23 July 2019 Page 1 of 20 Intersurgical Humidifier Heater Base PMH7000 Instructions for Use Distributed by Intersurgical Ltd Crane House Molly Millars Lane Wokingham Berkshire RG41 2RZ UNITED KING...

Page 2: ...late Temperature 7 3 1 8 Heater Wire Resistance 7 3 2 Physical Dimensions 7 3 2 1 Size Without Humidification Chamber Installed 7 3 2 2 Weight 7 3 3 Environmental 7 3 3 1 Operating Temperature 7 3 3 2...

Page 3: ...Left Side View 14 6 2 Left Side Panel Description 14 7 ASSEMBLY WITH BREATHING CIRCUIT 15 8 SET UP AND OPERATION OF THE HUMIDIFIER BASE 16 9 MAINTENANCE REPAIR 17 9 1 Basic Checks 17 9 2 Dual Tempera...

Page 4: ...vice 2 1 Warnings Cautions and Notes In this manual a WARNING describes the actions necessary to prevent injury or loss of life A CAUTION describes the actions to be taken to protect the equipment fro...

Page 5: ...n cause patient injury if not acted upon WARNING ALARMS No alarms sound or alarm location lights on the front panel can be activated while the system is shut down or while the protective thermostat is...

Page 6: ...ing Radiation Power On Off IPPV Mode NPPV Mode Alarm Silencer Button Do not discard 3 Specifications WARNING EXPLOSIVE HAZARD Do not use this equipment in the presence of flammable gases or anaestheti...

Page 7: ...10 to 95 Non condensing That includes the transportation time 3 3 5Atmospheric Pressure 700 to 1060 hPa 3 4 Electromagnetic Interference PMH7000 complies with limitations as specified in IEC 60601 1 2...

Page 8: ...y network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Guidance and manufacturer s declarat...

Page 9: ...ordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess electromagnetic environment due to fixed RF...

Page 10: ...0 Humidifier base incorporates a shunt fuse the PMH7000 is not compatible for use with electrical adaptors containing a shunt fuse 3 6 Settings for Ventilated Patients IPPV mode When the gas flow is i...

Page 11: ...est and alarm sound test as well When it is turned off alarm sound verification will be activated only once in short time Note In order to verify the functionality of the alarm system during the power...

Page 12: ...nk quickly in green and display patient temperature setting At this time temperature setting changes can be made by pressing up and down button Press the button once and the temperature value changes...

Page 13: ...rmittently LED Dual temperature probe Flashes red The dual temperature probe 7700000 is disconnected or faulty in this case a high priority audible alarm sounds intermittently LED Heater plate Flashes...

Page 14: ...Connector Temperature probe connecting port 3 Inspiratory Expiratory Heater Wire Connector Heated wire adaptor connecting port NOTE The power cord cannot be detached from the heater base The power co...

Page 15: ...g port Patient side temperature sensor Connect to the patient side temperature sensor connecting port on the breathing circuit Heater wire connection Inspiratory side Connect the connector of inspirat...

Page 16: ...connect the cloverleaf connector to inspiratory heated wire limb and oval connector to the expiratory heated limb When using a single heated wire systems connect the electrical adaptor 5600000 to the...

Page 17: ...n the dual temperature probe is exchanged or replaced be sure to turn off the power of the Humidifier base After exchange or replacement re activate the Humidifier base 12 Safe Process to stop operati...

Page 18: ...abnormal conditions during operation Each alarm shows the abnormal conditions by illuminating an LED on the front panel and sounding the alarm Alarm sound can be muted by the mute button for 2 minute...

Page 19: ...e inter lock is activated It is locked when alarm state continues for 10 minutes Turn off the device and restart it The device is not powered on No 1 Device fused 2 Heater base fault 3 Overheat protec...

Page 20: ...Instruction For Use PMH7000 Version 2 Revision 03 23 July 2019 Page 20 of 20...

Reviews: