Instruction For Use PMH7000
Version 2
Revision 03 23 July 2019
Page 15 of 20
7. Assembly with Breathing Circuit
As the preparation before use, the assembly process of breathing circuits is explained.
①
Refer to the breathing circuits assembling illustration, assemble breathing circuits.
②
Temperature probe or temperature probe connection
* Chamber outlet side temperature sensor: Connect to the chamber outlet side temperature probe connecting
port.
* Patient side temperature sensor: Connect to the patient side temperature sensor connecting port on the
breathing circuit.
③
Heater wire connection
* Inspiratory side: Connect the connector of inspiratory heater wire extension cord or inspiratory/expiratory heater
wire extension cord to the inspiratory breathing circuit.
* Expiratory side: When the dual heater wire breathing circuit is used, use the inspiratory/expiratory heater wire
extension cord (optional item). Connect the connector to expiratory breathing circuit.
“
Disposable Breathing Circuit Set-
Up”
With Single Heater Wire
“Disposable Breathing Circuit Set-Up”
With Dual Heater Wire