5 min

Disinfection of the breathing valve

Max 55°C 

  

!

A

55204A01

• Remove the cover from the 

breathing valve. Press on 

the threads to remove the 

diaphragm assembly. Re-

move the sealing disc and 

the spring with the guide 

disc. 

• Attach a breathing hose to 

the breathing valve. Make 

sure to not get any disin-

fectant into the open end of 

the hose. 

• Put the parts in the disin-

fectant and move it all 

around to get  

disinfectant on all surfaces. 

• Leave the parts in the 

disinfectant for 5 minu-

tes. Longer exposure time 

might cause damage to the 

equipment.

• Remove all parts from 

the disinfectant and 

pour out any remaining  

disinfectant. 

• Put all parts in fresh 

water and move it 

around to get rid of any 

disinfectant residues.

• Make sure that the thin, large 

O-ring on the diaphragm 

assembly still is in place. 

• Remove the breathing hose 

from the breathing valve.

•  Position the breathing valve 

with the openings in a position 

which makes it easier for the 

moisture to evaporate. 

•  The diaphragm assembly must 

not be flat during drying. Place 

it on an item (A) that keeps the 

diaphragm expanded to allow 

moisture to evaporate. 

•  Make sure there are no visible 

moisture on the breathing 

valve after drying. If necessary 

let it dry longer.

1

2

3

4

5

6

•  Dry all parts, maximum 55 °C for at least 2 h.

• Remove the parts 

from the water 

and shake off any 

remaining water.